Installation Summary
Страница 1: ...the suppressor comes from any tubing detector cell backpressure coils or any other components that are fluid connected to the eluent out port of the suppressor Step 1 Disconnect the column out eluent...
Страница 2: ...fore putting the new suppressor to use hydrate it properly first Improper hydration will result in irreversible damage to suppressor This will void the warranty Step 4 Connect the column out tubing to...
Страница 3: ...tep 5 Connect the suppressor Eluent outlet tubing to the detector cell inlet Then connect the detector cell outlet tubing with backpressure coil s to the Regen Inlet of the suppressor as shown in Figu...
Страница 4: ...t gases generated during AutoSuppression from out gassing by compressing the bubbles The bubbles can be occasionally trapped in the detector cell causing noise For example carbonate eluent is suppress...
Страница 5: ...open up the Suppressor tab ICS 6000 or Electrolytics tab Integrion Double click on the Suppressor option and click OK to save the new configuration see Figures 3 4 below If this is not done CM will no...