Troubleshooting Guide
7. Calibrate
the
grating
a. Move the grating to 10 steps below the peak.
b. Scan to 10 steps above the peak.
c. Find the actual energy peak in the scan.
d. On a VIS instrument, repeat steps 7.a – 7.c for 4 wavelengths
i. 494.0nm
peak
ii. 585.2nm
valley
iii. 775.0nm
peak
iv. 807.1nm
valley
e. On the UV instrument, repeat steps 7.a – 7.c for 4 wavelengths
i. 824.1nm
peak
ii. 882.7nm
peak
iii. 895.8nm
peak
iv. 905.2nm
peak
v. 917.0nm
peak
f. Set the final calibration by the average of the errors from the calibration peaks.
8.
On the VIS instrument, move to 681nm and determine the ADC integration time.
9.
Move to the last wavelength used on the walkup (ATC) screen, and zero the system.
Spectronic GENESYS 10 UVscanning, Spectronic GENESYS 10 > Revision 1.103, Spectronic
BioMate 3 > Revision 1.001 and Spectronic GENESYS 10 Bio > Revision 1.000 Initialization
1. Power
on.
2. Calibrate
turret
a. Rotate the turret while reading the turret sensor.
b. After we have detected 3 small openings without finding the big opening, reverse the
turret rotation direction.
c. After we have detected 6 more small openings without finding the big opening, return a
turret calibration failure error.
d. When we find the big opening (indicates position B), center the turret in the opening.
3.
Initialize the grating
a. Moving to the grating calibration switch.
b. On the VIS instrument, move to grating step 8695 (ideally 1095nm).
c. On the Spectronic GENESYS 10 UVscanning, move to grating step 15948 (ideally
716nm).
d. On other UV instruments, move to grating step 6475 (ideally 850nm).
4.
Initialize the filter wheel
a. Rotate the filter wheel while recording energy levels.
b. Look for an area that passes light.
c. Center the filter wheel in the position found.
5.
Find zero order
a. Spectronic GENESYS 10 UVscanning.
i. Scan between grating steps 1472 and 2432 for the start of zero order saturation.
ii. Scan towards grating step 3968 for the end of zero order saturation.
b. Spectronic GENESYS 10, Spectronic BioMate 3 and Spectronic GENESYS 10 Bio.
i. Scan between grating steps 555 and 2035 for the start of zero order saturation.
ii. Scan towards grating step 2405 for the end of zero order saturation.
c. Set the initial calibration to the center of the zero order area.
3-2
Содержание Spectronic BioMate 3
Страница 1: ......
Страница 6: ...This page intentionally left blank vi ...
Страница 13: ...Optical Diagram Figure 1 1 Optical Diagram Shown are both UV Vis and Vis Optics 1 5 ...
Страница 14: ...Specifications This page intentionally left blank 1 6 ...
Страница 22: ...Troubleshooting Guide This page intentionally left blank 3 6 ...
Страница 36: ...Routine Maintenance This page intentionally left blank 4 8 ...
Страница 72: ......
Страница 78: ...Drawings This page intentionally left blank 8 2 ...
Страница 79: ...Drawings 8 3 Figure 8 1 Spectronic GENESYS 10 Wiring Diagram ...
Страница 80: ...wings 8 4 This page intentionally left blank Dra ...
Страница 81: ...Drawings 8 5 Figure 8 2 Keypad Schematic ...
Страница 82: ...wings 8 6 This page intentionally left blank Dra ...
Страница 84: ...wings 8 8 This page intentionally left blank Dra ...
Страница 86: ...Drawings This page intentionally left blank 8 10 ...
Страница 88: ...Drawings This page intentionally left blank 8 12 ...
Страница 90: ...Drawings This page intentionally left blank 8 14 ...
Страница 92: ...Drawings This page intentionally left blank 8 16 ...
Страница 94: ...Drawings This page intentionally left blank 8 18 ...
Страница 96: ...Drawings This page intentionally left blank 8 20 ...
Страница 98: ...Drawings This page intentionally left blank 8 22 ...
Страница 100: ...Drawings This page intentionally left blank 8 24 ...
Страница 102: ...Drawings This page intentionally left blank 8 26 ...
Страница 104: ...Drawings This page intentionally left blank 8 28 ...
Страница 106: ...Drawings This page intentionally left blank 8 30 ...
Страница 108: ...Drawings This page intentionally left blank 8 32 ...
Страница 110: ...Drawings This page intentionally left blank 8 34 ...
Страница 112: ...Drawings This page intentionally left blank 8 36 ...
Страница 114: ...Drawings This page intentionally left blank 8 38 ...
Страница 116: ...Drawings This page intentionally left blank 8 40 ...
Страница 118: ...Drawings This page intentionally left blank 8 42 ...
Страница 120: ...Drawings This page intentionally left blank 8 44 ...
Страница 122: ...Drawings This page intentionally left blank 8 46 ...
Страница 124: ...Drawings This page intentionally left blank 8 48 ...
Страница 126: ...Drawings This page intentionally left blank 8 50 ...
Страница 128: ...Drawings This page intentionally left blank 8 52 ...
Страница 130: ...Drawings This page intentionally left blank 8 54 ...
Страница 132: ...Drawings This page intentionally left blank 8 56 ...
Страница 134: ...Drawings This page intentionally left blank 8 58 ...
Страница 136: ...Drawings This page intentionally left blank 8 60 ...
Страница 138: ...Drawings This page intentionally left blank 8 62 ...
Страница 140: ...Drawings This page intentionally left blank 8 64 ...
Страница 142: ...Drawings This page intentionally left blank 8 66 ...
Страница 144: ...Drawings This page intentionally left blank 8 68 ...
Страница 146: ...Drawings This page intentionally left blank 8 70 ...
Страница 148: ...Drawings This page intentionally left blank 8 72 ...
Страница 150: ...Drawings This page intentionally left blank 8 74 ...
Страница 152: ...Drawings This page intentionally left blank 8 76 ...
Страница 166: ...Service Diagnostic Software This page intentionally left blank 9 14 ...
Страница 170: ...Technical Support Bulletins This page intentionally left blank 10 4 ...