19
..... Additional setup menus within the TOTAL submenu
F
IGURES
3.11-E (
LEFT
) & 3.11-F (
RIGHT
)
3.12 Additional Setup Menus Within the OPTIONS Submenu
*Not included in the Quick Setup configuration
The
OPTIONS
submenu contains setup menus for several miscellaneous functions.
1.
Measurement Units
setup menu (42): Use this menu to select
ENGLISH
(feet per second,
FPS
)
or
METRIC
(meters per second,
MPS
) measurement units (Figure 3.12-A, below).
2.
Site Parameters
setup menu (43): This menu saves the parameters for the pipe, liner, fluid,
transducer, and flow setup menus, allowing them to be recalled later for a specific measurement site.
Several sites are available and are numbered. The site number is displayed in Menu 43 in the lower
left corner of the screen and is followed by a colon. See Figure 3.12-B (below) for an example of a site
numbered 1.
As you enter the setup parameters in their respective menus during normal configuration, the same
parameters are saved simultaneously in the
Site Parameters
setup menu, in whichever site has the
asterisk displayed.
Access a different site to automatically enter that site’s stored parameters into the flowmeter for
measuring flow. To access a different site, press
ENTER
, scroll to the desired site, and press
ENTER
again.
To avoid overwriting stored parameters and losing old data, ensure that the desired site
is active prior to entering the new set of parameters.
F
IGURES
3.12-A (
LEFT
) & 3.12-B (
RIGHT
)