7
•
Maintenance
Vanquish Charger Operating Manual
Page 61
NOTICE
A firmware downgrade or incomplete firmware update may
result in loss of functionality or malfunctioning of the Charger.
•
Do not interrupt communication between Chromeleon and the
Charger at any time during the procedure.
•
At the beginning of the update process, a message appears showing
the firmware version currently installed in the Charger and the
version that will be transferred from Chromeleon. If the firmware
installed in the Charger is a newer version than the version in
Chromeleon, cancel the download.
The firmware update for the Charger usually takes longer than a
firmware update for the other Vanquish modules, usually up to a few
minutes.
•
Monitor the Audit Trail of the Chromeleon Instrument Configuration
Manager (or Server Configuration program) to find out if the
firmware update was successful or failed.
•
If the firmware update failed, turn the device off and on again and
repeat the firmware update.
•
If the firmware update fails repeatedly, contact Thermo Fisher
Scientific Technical Support for assistance.
Содержание Vanquish Charger
Страница 1: ...Vanquish Charger Operating Manual 4829 0001 EN Revision 1 0a September 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Charger Operating Manual ...
Страница 8: ...Contents Page 8 Vanquish Charger Operating Manual ...
Страница 66: ...7 Maintenance Page 66 Vanquish Charger Operating Manual ...
Страница 76: ...9 Specifications Page 76 Vanquish Charger Operating Manual ...
Страница 80: ...10 Accessories Consumables and Replacement Parts Page 80 Vanquish Charger Operating Manual ...
Страница 85: ...12 Index Vanquish Charger Operating Manual Page 85 12 Index ...
Страница 88: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...