Process
OQ/PQ Operating Instructions
M1393R0
Page 17 of 60
•
FAMOS with Cooling Option:
;Commands for FAMOS with cooling option
Sampler.TempCtrl =
On
Sampler.Temperature.Nominal =
20.00
Sampler.Temperature.LowerLimit
=
5.00
Sampler.Temperature.UpperLimit
=
30.00
•
FAMOS without Cooling Option:
;Commands for FAMOS with cooling option
;Sampler.TempCtrl =
On
;Sampler.Temperature.Nominal =
20.00
;Sampler.Temperature.LowerLimit = 5.00
;Sampler.Temperature.UpperLimit =
30.00
3.2.1.B.4
Oven Support
During the warm up sequence of the UV detector noise and drift test (Section 4.6) the
oven accuracy is checked. If the current configuration does not include an oven, a different
program must be used:
Depending on the test configuration (e.g. the Switchos is included or not),
•
If the system includes an oven, run the ‘warm up and oven test.pgm’ (FIGURE 3-3).
•
If the system does not include an oven, run the ‘warm up.pgm’ (FIGURE 3-3).
FIGURE 3-3
Selecting the proper Warm-up Program
3.2.1.B.5
Miscellaneous
The sequences for the test 4.10 and 4.12 include a stop flow method which reduces the
time that the lamp is lit and minimizes solvent consumption. Select the status of these
stop flow samples to ‘single’ (activated) or ‘interrupted’ (deactivated) to correspond to the
test configuration (e.g. the Switchos is included or not).
3.2.1 C
Report File
No changes should be made in the report file itself (PQ_OQ_LCP.rdf), except for the
spreadsheets ‘Specification’ and ‘Others tests’. These two spreadsheets allow entering of
some test details (i.e. the serial numbers of the instruments) and the results of tests that
were carried out manually (i.e. 4.2, 4.4, 4.5, 4.11, 4.13).