- 8 -
Electrical
Requirements
The circulator construction provides protection against the risk of
electrical shock by grounding appropriate metal parts. The protection
will not function unless the power cord is connected to a properly
grounded outlet. It is the user's responsibility to assure a proper
ground connection is provided.
Make sure the voltage of the power source meets the specified voltage, ±10%.
Minimum Circuit Ampacity
ULT-80 50/60 Hz
16A
ULT-95 50/60 Hz
20A
+80°C Circulator 50/60Hz
21A
Plug 50/60 Hz
L6-30P
Refer to the serial number label on the rear of the circulator to identify the
specific electrical requirements.