4
Maintenance
Servicing the LDA
Thermo Scientific
Surveyor MS Pump and Surveyor MS Pump Plus Hardware Manual
45
Figure 37.
Loosening the two fastening screws
To provide enough clearance to remove the piston guide bushings, loosen the screws by at
least six turns. As you loosen the screws, the holding hooks move to the right. If you
prefer, you can detach the holding hooks completely with approximately nine turns on
the screw fasteners. See
.
Figure 38.
View of loosened holding hooks
Fastening screws
Holding hooks