Installation
Connecting the Low-Pressure Solvent Lines
Thermo Scientific
Surveyor MS Pump and Surveyor MS Pump Plus Hardware Manual
15
Figure 12.
Back panels, showing the connections for the 7-connector interconnect cable
Connecting the Low-Pressure Solvent Lines
You need to connect the low-pressure solvent lines to the Surveyor MS Pump Plus before you
operate the pump. There are four low-pressure lines that bring the four solvents to the pump.
Solvents should always be HPLC quality and free of particulate matter. Because the Surveyor
MS Pump Plus contains very small, precision-machined parts, it can easily be affected by
small particles entering the pump. Aqueous eluents are particularly susceptible to bacterial
growth that can cause numerous problems.
To prevent bacteria and/or particulate matter from entering the pump, do the following:
• Filter all solvents thoroughly.
• Discard aqueous mobile phases on a daily basis or add a bactericide such as sodium azide
to inhibit bacterial growth (0.65 g/l NaN
3
prevents growth of most kinds of bacteria).
DETECT
OR
DETECT
OR
PUMP
MS PUMP
Solvent platform
Autosampler
(back panel)
Surveyor
MS Pump
(back panel)
Adapter cable for MS pump
7 inches (18 cm) P/N 60053-63038
Detector
(back panel)
To
Pump connector
8-pin
connector
CAUTION
Sodium azide is highly toxic and carcinogenic. Take care when disposing of
solutions that contain sodium azide because azides can react with heavy metals to form
explosive compounds.