Installation
Making the Initial Back Panel Connections
12
Surveyor MS Pump and Surveyor MS Pump Plus Hardware Manual
Thermo Scientific
Serial Communication Cable
The serial communication cable is a standard RS232 cable, approximately 2 meters long and
terminated at each end with a DB-9 pin connector. The serial communication cable is in the
accessory kit.
To install the serial communication cable
1. Connect one end of the serial communication cable to an available serial port on the host
computer.
2. Connect the other end of the cable to the connector found on the back of the Surveyor
MS Pump Plus (see
System Interconnect Cable
The interconnect cable coordinates the timing of the Surveyor Plus modules during an
injection sequence.
The Surveyor Plus LC devices are shipped with two versions of the interconnect cable: a
5-connector version (F5049-010) and a 7-connector version (60053-63034).
This topic contains the following procedures:
•
Connecting the 5-Connector Interconnect Cable to the LC Devices
•
Connecting the 7-Connector Interconnect Cable to the LC Devices
Connecting the 5-Connector Interconnect Cable to the LC Devices
The 5-combicon connectors of the interconnect cable (P/N F5049-010) are labeled LC
PUMP, MS PUMP, A/S, DET, and M/S. See
.
Figure 8.
5-connector interconnect cable, P/N F5049-010
LC PUMP
MS PUMP
M/S
A/S
DET
F5049-010
To
MS detector
To
MS pump
To
Autosampler
To
detector
To
LC pump