UltiMate 3000 Series:
SRD-3x00 Solvent Racks
Page
26
Operating Instructions
When replacing a solvent supply line, remove the filter holder, then the retaining guide,
and then the solvent supply line.
7.
Connect the lines to the appropriate connection ports on the degassing module. Route
the solvent supply lines through the opening between the housing and the front panel
door.
Fig. 8: Opening for the solvent supply lines
8.
Remove the fitting plugs from the degasser inlet if necessary and connect the solvent
supply line with the associated line label.
Fig. 9: Example for degasser connections (here SRD-3600)
Tip:
The degasser ports in the upper row are combined in pairs with the related
ports in the lower row.
Before connecting the solvent supply lines, make sure that the connectors are free of
contaminants. Even minute particles can allow air to enter the degasser, and thus reduce
the degassing effectiveness.
9.
Connect the solvent supply lines from the pump to the degasser (
→
page 27).
A B C A
B C
Degasser inlets
(solvent supply line
Degasser outlets
(to pump)
Tubing guides
Содержание SRD-3200
Страница 2: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page II Operating Instructions ...
Страница 28: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 22 Operating Instructions ...
Страница 34: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 28 Operating Instructions ...
Страница 44: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 38 Operating Instructions ...
Страница 54: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 48 Operating Instructions ...
Страница 56: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 50 Operating Instructions ...
Страница 60: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 54 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 56 Operating Instructions ...