UltiMate 3000 Series:
SRD-3x00 Solvent Racks
Operating Instructions
Page
43
Problem
Probable Cause
Remedial Action
Spikes
There is electrical interference
from other modules.
Isolate the electrical circuit
from strong current consumers.
Consider installing an
uninterruptible power supply
(UPS).
For additional causes, refer to the
operating instructions for the
detector and for the column
compartment.
→
Detector manual
→
TCC manual
Negative peaks
Sample solvent and mobile phase
differ in composition.
Dissolve the sample in the
mobile phase.
The absorption of the solute is
lower than the absorption of mobile
phase.
Select a different wavelength.
Use a mobile phase with less
UV background absorption.
Poor peak area precision
The capillary connections are not
installed properly or they are not
tight.
Check and tighten the capillary
connections. Consider using
Viper capillaries.
Exchange the needle seat if
necessary (
→
Autosampler
Manual
).
Exchange the needle if
necessary (
→
Autosampler
manual
).
There are dead volumes in the
capillary connections.
Replace the fittings. Make sure
that the capillaries are installed
correctly. Consider using Viper
capillaries.
The sample is unstable and
decomposes.
Use new sample or change the
conditions. Cool the sample in
the autosampler if possible.
Baseline fluctuations
See the remedial actions
provided in the related baseline
sections further up in this table.
Содержание SRD-3200
Страница 2: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page II Operating Instructions ...
Страница 28: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 22 Operating Instructions ...
Страница 34: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 28 Operating Instructions ...
Страница 44: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 38 Operating Instructions ...
Страница 54: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 48 Operating Instructions ...
Страница 56: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 50 Operating Instructions ...
Страница 60: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 54 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 56 Operating Instructions ...