1
Introduction and Description
Technical Data
Thermo Scientific
ST16 / 16R
1-3
Technical Data
The technical data of the Sorvall ST 16/16R is listed in the following table.
Table 1-1.
Technical Data Sorvall ST 16/16R
Feature
Value
Environmental conditions
-Use in interior spaces
-Altitudes of up to 2,000 m above sea level
-max. relative humidity 80 % up to 31 °C; decreasing linearly up to 50 %
relative humidity at 40 °C.
permissible ambient temperature
+2 °C to +35 °C
+2 °C to +35 °C
Overvoltage category
II
II
Pollution degree
2
2
Heat dissipation
refrigerated
ventilated
230V
120V
230V
120V
4778 BTU/h
4096 BTU/h
3447 BTU/h
3901 BTU/h
IP
20
20
running time
unlimited
unlimited
max Speed n
max
15200 rpm (depending on the rotor)
15200 rpm (depending on the rotor)
min Speed n
min
300 rpm
300 rpm
maximum RCF value at n
max
25830xg 25830xg
maximum kinetic energy
62.5 kJ
51.7 kJ
noise level at maximum speed
< 63 dB (A)
< 63 dB (A)
temperature setting range
-10 °C to +40 °C
Dimensions
refrigerated
ventilated
Height
Height with lid open
Width
Depth
Table top height
360 mm
870 mm
623 mm
605 mm
310 mm
360 mm
870 mm
440 mm
605 mm
310 mm
weight without rotor
91.5 kg
57.5 kg