Use the optional cooling coil to maintain bath temperature at or below
ambient temperature.
1. Remove rear splash baffle.
2. Mount cooling coil on 2 clips at the back of the baffle on 6" centers.
3. Centerline of coil should be 3/4" (1.9cm) above top of baffle. Crimp
the clip tabs together using pliers.
4. Remove knobs that clamp suspension and support arms together.
Remove shaking platform from tank.
5. Install the rear baffle and cooling coil on the back of tank. Cooling coil
is shaped to fit around the heater. Two compression fittings
accommodate 5/16" (0.79cm) OD inlet and outlet tubing (supplied by
user).
6. Re-install suspension arms and platform.
Note
Shipped cooling coils is compressed to 3-4” across. Coil needs to be
expanded to 6” for installation. This does not affect coil performance as
copper coil is annealed for flexibility.
4-4
MaxQ 7000
Thermo Scientific
Section 4
Unpacking and Installation
Optional Cooling Coil
Installation
6"
(15.3 cm)
Crimp here
Crimp here
Baffle Plate
3/4" (1.9 cm)
Cooling Coil
Crimp here
Crimp here
Figure 4-2.
Clips
Figure 4-3.
Crimp Locations