2-2
MaxQ 7000
Thermo Scientific
Section 2
Specifications
Model number
SHKA7000
SHKA7000-1CE
SHKE7000
SHKE7000-1CE
Speed Accuracy
40 to 400 rpm, ±10rpm
40 to 400 rpm, ±10rpm
15 to 500 rpm, ±1rpm
15 to 500 rpm, ±1rpm
Timer
Continuous or timed opera-
tion from 1-60 min.
Continuous or timed opera-
tion from 1-60 min.
Continuous or timed opera-
tion 0.1-999 hours or 0.1-999
mins.
Continuous or timed opera-
tion 0.1-999 hours or 0.1-999
mins.
Display
3 individual LED displays
indicate temperature in 1°C
increments. Analog tachome-
ter displays speed in rpm.
3 individual LED displays
indicate temperature in 1°C
increments. Analog tachome-
ter displays speed in rpm.
3 individual LED displays
indicate temperature, time
and speed simultaneously. 3
characters height ½ inch
(1.27 cm)
3 individual LED displays
indicate temperature, time
and speed simultaneously. 3
characters height ½ inch
(1.27 cm)
Mutable Alarms
None
None
Audible portion of the alarm
can be silenced for a period
of 1 hour without deactivat-
ing the actual alarm condi-
tion by depressing any key.
Audible portion of the alarm
can be silenced for a period
of 1 hour without deactivat-
ing the actual alarm condi-
tion by depressing any key.
Motor
Permanent Magnet DC
Permanent Magnet DC
Solid State Brushless DC
Solid State Brushless DC
Soft Start Feature
None
None
Software algorithms prevent
sudden start/stops.
Software algorithms prevent
sudden start/stops.
RS232 Interface *
None
None
Monitor speed, temperature
in °C and time with a com-
puter.
Monitor speed, temperature
in °C and time with a com-
puter.
Recorder Output *
None
None
10 mv/°C output monitors
temperature with external
chart recorder.
10 mv/°C output monitors
temperature with external
chart recorder.