SYMPTOM
REFRIGERATED CONDENSATION
TRAP
Refrigerated trap not cooling.
Noisy vibration and/or excessive heat
on top of cabinet.
Glass Condensation Flask breakage.
Glass Condensation Flask plugged with
ice.
Vacuum leak at flask cap.
POSSIBLE CAUSES
Insufficient air circulation.
Compressor system has lost it charge.
Cabinet housing has become loose or
circulation fan is rubbing an internal
component.
Large ice build-up that expands during
thawing.
Trap contents allowed to thaw and
refreeze.
Excessive water/ice build up in thermal
transfer fluid.
Flask Cap seated improperly.
Flask cap worn or fouled.
SOLUTION/EXPLANATION
Leave a minimum four-inch clearance
on all sides. Clean the condenser of
any accumulated dust or debris.
Call Thermo Scientifice for assistance.
Call Thermo Scientific for assistance.
Change and clean Glass Condensation
Flask daily.
Leave the main power switch ON
between runs and change the Glass
Condensation Flask daily.
Use CryoCool
®
instead of ethanol.
Remove the ice. CAUTION:
Extreme
cold may cause severe blistering.
Verify that the cap is pressed securely
over
the
mouth
of
the
Glass
Condensation Flask.
Replace flask cap.