SPEEDVAC

®

CHAMBER

Rotor does not rotate or makes a grind-
ing noise.

Rotor makes a loud clattering noise on
initial start up.

Chamber does not seal properly.

Chamber does not reach temperature.

No current to the SpeedVac

®

Badly

corroded

upper

magnetic

assembly.

Lower drive magnet bound up on the
separation plate.

Hold-down knob has been over-tight-
ened.

Cracked or fouled cover gasket.

Automatic bleeder valve malfunction.

Radiant lamp burned out.

Heater control or element malfunction.

Check line cord.

Upper magnetic assembly requires
replacement. Call Thermo for service.

Call Thermo for Service.

Tighten hold-down knob until it make
contact with rotor,
DO NOT OVER-TIGHTEN.

Clean or replace cover gasket.

Call Thermo for assistance.

Replace lamp, or call Thermofor
assistance.

Call Thermo for assistance.

7

SYMPTOM

GENERAL

No Response.

System trips circuit breaker
or main line fuse.

Sample does not dry or noticeable
change in the drying rate.

POSSIBLE CAUSES

No current to the equipment.
Power failure at source.

Short circuit to ground in one of the
components.

Flask cap plugged with ice.

Internal vacuum pump needs servic-
ing.

Vacuum leak.

Insufficient thermal energy applied to
the sample chamber.

Refrigerated trap not reaching operat-
ing temperature.

Condensation flask full past 3/4 mark.

SOLUTION/EXPLANATION

Check line cord for proper connections.
Check main circuit panel.

Call Thermo for assistance.

Remove blockage. Change to dry flask.

Call Thermo for assistance.

Check

seal

of

cap

on

Glass

Condensation Flask. Check cover gas-
ket on sample chamber.

Evaporative cooling retards evapora-
tion; additional chamber heat will
counteract this effect.

Call Thermo for assistance.

Change to new dry flask.

TROUBLESHOOTING GUIDE

ISS110 SPEEDVAC

®

TROUBLESHOOTING GUIDE

Содержание Savant Speedvac ISS110

Страница 1: ...Instruction Manual Thermo Scientific Savant ISS110 Speedvac Concentrator ...

Страница 2: ......

Страница 3: ... 2008 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Страница 4: ...er Cover 2 3 4 Simple System IntegrityTest 3 Section Application Information Page 4 0 Application Information 3 4 1 Solvent Handling 3 4 4 2 Accessories 4 4 3 Devising Protocols 4 Section Accessories Page 5 0 Accessories 5 Section TroubleShooting Page 6 0 Troubleshooting 5 Troubleshooting Guide 7 8 Section Specifications Page 7 0 Specifications 5 Section Return Page 8 0 To Return an Instrument for...

Страница 5: ...es a stable level surface for proper operation Units configured for 115 VAC 60Hz should be plugged into a circuit rated for at least 10 amps Units configured for 230 VAC 50Hz should be plugged into a circuit rated for at least 5 amps IMPORTANT INFORMATION Before operating the unit read section 3 0 Operation and Section 4 0 Application Information to determine your specific application requirements...

Страница 6: ...he lid Always balance rotor loads An unbalanced rotor causes vibration that will damage the bearings Load the rotor symmetrically There need not be a tube in each holder but you must be sure to evenly balance the rotor by placing an even number of tubes opposite each other When using a rotor with aluminum tube holders insert all tube holders By using the Drying Rate switch thermal energy can be ap...

Страница 7: ...y of the vacuum pump and refrigerated vapor trap 1 Install a clean dry glass condensation flask 2 Allow the refrigerated vapor trap to operate for 45 minutes 3 Open the chamber lid and remove the rotor 4 Fill a plastic beaker that will fit in the chamber with 50 ml of water and place it in the chamber Close the lid 5 Press the drying rate switch to the LOW position 6 Start the system by pressing t...

Страница 8: ...121 Ammonia Neutralizing Solution bottles which screw onto the ANT100 neutralize ammonia gas preventing unpleasant odors in the laboratory The CC120 DX Deluxe Convenience Cart is a useful accessory for the envi ronmental lab The SpeedVac resides on the top shelf while any additional trap ping apparatus are placed on the bottom shelf Addition of the cart produces a completely mobile concentration s...

Страница 9: ...ion resistant fully adjustable polypropylene shelves and swivel casters DVG50 Digital Vacuum Gauge Ideal for monitoring the vacuum efficiency or sample drying efficiency Insertion into the vacuum line between the rotor cham ber and the refrigerated trap provides vapor pressure readings which can be cor related to sample dryness level Insertion into the vacuum line between the refrig erated trap an...

Страница 10: ...ceive 1 an Return material Authorization RMA number and 2 sign the Health and Safety Clearance Form before shipping the equipment back All items returned must be certified to be decontaminated and free of radioactivity When returning equipment that may contain hazardous materials you must pack and label them in accordance with DOT regulations applying to the transportation of hazardous material Yo...

Страница 11: ...hermo for assistance 7 SYMPTOM GENERAL No Response System trips circuit breaker or main line fuse Sample does not dry or noticeable change in the drying rate POSSIBLE CAUSES No current to the equipment Power failure at source Short circuit to ground in one of the components Flask cap plugged with ice Internal vacuum pump needs servic ing Vacuum leak Insufficient thermal energy applied to the sampl...

Страница 12: ... thaw and refreeze Excessive water ice build up in thermal transfer fluid Flask Cap seated improperly Flask cap worn or fouled SOLUTION EXPLANATION Leave a minimum four inch clearance on all sides Clean the condenser of any accumulated dust or debris CallThermo Scientifice for assistance CallThermo Scientific for assistance Change and clean Glass Condensation Flask daily Leave the main power switc...

Страница 13: ...er www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodotto verrà smaltito o riciclato tramite ...

Страница 14: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...

Страница 15: ......

Страница 16: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 134 3001 00 Rev C ...

Отзывы: