2. Gently put a clean Glass Condensation Flask (GCF400) into the refrigerated
chamber. As you lower the flask into the chamber, the CryoCool Fluid rises.
Verify that the final CryoCool liquid level is 10 to 15 mm below the shoulder of
the flask.

3. Immediately wipe clean any CryoCool Fluid that spills onto the rubber seal.

4. Fit the white insulating Flask Seal over the glass flask to secure the flask in the
chamber. Its beveled side faces upward to admit the Flask Cap.

5. Snap the black rubber Flask Cap over the mouth of the glass flask. This pro-
vides easy tubing connection and a vacuum seal while also securing the Glass
Condensation Flask and insulating Flask Seal in the refrigerated chamber.

Rotor Installation

. Open the lid of the rotor chamber. Visually align the pin on

the drive shaft with the groove on the bottom of the rotor. Carefully lower the
rotor onto the drive shaft. Rotate the rotor by hand to be sure the pin is lined up
with the groove. Secure the assembly by screwing the retaining knob into the
drive shaft above the rotor. Tighten it firmly but not excessively. Load the rotor
with samples and close the lid.

Always balance rotor loads. An unbalanced rotor causes vibration that will
damage the bearings. Load the rotor symmetrically. There need not be a tube in
each holder, but, you must be sure to evenly balance the rotor by placing an
even number of tubes opposite each other. When using a rotor with aluminum
tube holders, insert all tube holders.

By using the Drying Rate switch, thermal energy can be applied to the
concentrator chamber to counteract the cooling effect of evaporation, maintain
the samples in the liquid state, and accelerate the concentration run.
The DRYING RATE switch selects the desired drying rate:

LOW maintains the concentrator chamber at ambient temperature.

MEDIUM maintains the concentrator chamber at approximately 43

°

C.

HIGH maintains the concentrator chamber at approximately 65

°

C.

Select the drying rate according to the nature of your samples.

The CONCENTRATOR switch is normally in the OFF position. After loading and
installing the rotor, closing the lid, selecting a DRYING RATE (you may even
pre-heat the samples), press the CONCENTRATOR switch to the ON position to
start the drying run. To stop the run, press the CONCENTRATOR switch to the
OFF position. A run can be stopped at any point and restarted based on your
criterion for sample dryness.

3.1

DRYING RATE

3.2

STARTING AND STOPPING A RUN

2

Содержание Savant Speedvac ISS110

Страница 1: ...Instruction Manual Thermo Scientific Savant ISS110 Speedvac Concentrator ...

Страница 2: ......

Страница 3: ... 2008 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Страница 4: ...er Cover 2 3 4 Simple System IntegrityTest 3 Section Application Information Page 4 0 Application Information 3 4 1 Solvent Handling 3 4 4 2 Accessories 4 4 3 Devising Protocols 4 Section Accessories Page 5 0 Accessories 5 Section TroubleShooting Page 6 0 Troubleshooting 5 Troubleshooting Guide 7 8 Section Specifications Page 7 0 Specifications 5 Section Return Page 8 0 To Return an Instrument for...

Страница 5: ...es a stable level surface for proper operation Units configured for 115 VAC 60Hz should be plugged into a circuit rated for at least 10 amps Units configured for 230 VAC 50Hz should be plugged into a circuit rated for at least 5 amps IMPORTANT INFORMATION Before operating the unit read section 3 0 Operation and Section 4 0 Application Information to determine your specific application requirements...

Страница 6: ...he lid Always balance rotor loads An unbalanced rotor causes vibration that will damage the bearings Load the rotor symmetrically There need not be a tube in each holder but you must be sure to evenly balance the rotor by placing an even number of tubes opposite each other When using a rotor with aluminum tube holders insert all tube holders By using the Drying Rate switch thermal energy can be ap...

Страница 7: ...y of the vacuum pump and refrigerated vapor trap 1 Install a clean dry glass condensation flask 2 Allow the refrigerated vapor trap to operate for 45 minutes 3 Open the chamber lid and remove the rotor 4 Fill a plastic beaker that will fit in the chamber with 50 ml of water and place it in the chamber Close the lid 5 Press the drying rate switch to the LOW position 6 Start the system by pressing t...

Страница 8: ...121 Ammonia Neutralizing Solution bottles which screw onto the ANT100 neutralize ammonia gas preventing unpleasant odors in the laboratory The CC120 DX Deluxe Convenience Cart is a useful accessory for the envi ronmental lab The SpeedVac resides on the top shelf while any additional trap ping apparatus are placed on the bottom shelf Addition of the cart produces a completely mobile concentration s...

Страница 9: ...ion resistant fully adjustable polypropylene shelves and swivel casters DVG50 Digital Vacuum Gauge Ideal for monitoring the vacuum efficiency or sample drying efficiency Insertion into the vacuum line between the rotor cham ber and the refrigerated trap provides vapor pressure readings which can be cor related to sample dryness level Insertion into the vacuum line between the refrig erated trap an...

Страница 10: ...ceive 1 an Return material Authorization RMA number and 2 sign the Health and Safety Clearance Form before shipping the equipment back All items returned must be certified to be decontaminated and free of radioactivity When returning equipment that may contain hazardous materials you must pack and label them in accordance with DOT regulations applying to the transportation of hazardous material Yo...

Страница 11: ...hermo for assistance 7 SYMPTOM GENERAL No Response System trips circuit breaker or main line fuse Sample does not dry or noticeable change in the drying rate POSSIBLE CAUSES No current to the equipment Power failure at source Short circuit to ground in one of the components Flask cap plugged with ice Internal vacuum pump needs servic ing Vacuum leak Insufficient thermal energy applied to the sampl...

Страница 12: ... thaw and refreeze Excessive water ice build up in thermal transfer fluid Flask Cap seated improperly Flask cap worn or fouled SOLUTION EXPLANATION Leave a minimum four inch clearance on all sides Clean the condenser of any accumulated dust or debris CallThermo Scientifice for assistance CallThermo Scientific for assistance Change and clean Glass Condensation Flask daily Leave the main power switc...

Страница 13: ...er www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodotto verrà smaltito o riciclato tramite ...

Страница 14: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...

Страница 15: ......

Страница 16: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 134 3001 00 Rev C ...

Отзывы: