Chapter 2
Thermo Fisher Scientific
User Guide for Sarasota ID900, FD900, & PD900 Gas Density Meters
2-1
Installation
Note
Installation must be carried out in accordance with local site
requirements and regulations.
▲
Sarasota ID900
The Sarasota ID900 should be mounted vertically in the pipeline with the
outlet port downstream. There are three mountingmethods:
●
Welded flange version
●
Male connector Swagelok fitting
●
Sarasota RTR900 instrument retractor
Ensure that the flow direction arrow etched onto the Sarasota ID900 stems
points in the flow stream direction. Fit suitable flange gaskets and tighten
flange bolts evenly. See the figures below.