background image

 
 
 
 

 
 

 

 

 

Pierce Biotechnology 

PO Box 117 

(815) 968-0747 

www.thermo.com/pierce

 

 

 

 

3747 N. Meridian Road 

Rockford, lL 61105 USA 

(815) 968-7316 fax 

 

 

4

Note:

 Once calibration has been performed for a given aluminum block, vial size and ambient condition, setting the 

Heat knob to the same set temperature with the Display Adjust knob at the same dial position for subsequent uses 

will reproduce the same heating performance. You may wish to mark or note the Display Adjust dial position for 
future reference when using this same heating application.  

 

Instructions for Use with Remote Temperature Probe (RTP, Product No. 18820) 

Using the Remote Temperature Probe (RTP) provides more accurate and responsive control of the heating function because it 
replaces the internal heat sensor (housed under the block reservoir) with one placed directly in the aluminum block or 
reaction vial. However, the RTP is not a substitute for temperature calibration to an independent thermometer. 

For best results, perform temperature calibration for a given application (i.e., particular temperature setting, aluminum block 
and vial/sample size) with the RTP placed in the aluminum block thermometer well and a standard thermometer placed in a 
reaction vial in one of the aluminum block wells. Calibration performed in this configuration will provide for accurate in-vial 
sample heating, controlled by the RTP conveniently positioned in the thermometer well. The triple-block module allows for 
both the RTP and thermometer to be placed in block thermometer wells, making alternative configurations possible.  
Calibrate with the RTP in a vial only if that will be the configuration for use. The following instructions assume the 
suggested calibration configuration. Be aware that the aluminum block and in-vial sample temperatures may be different. 

1.

 

Place clean Reacti-Block Aluminum Block(s) into the module reservoir. 

Note:

 The triple-block reservoir space of the 

Reacti-Therm III Module must be filled with three aluminum blocks to obtain acceptable heating performance. 

2.

 

Set Heat knob to the off position (turn all the way counter-clockwise). Set the Display Adjust knob to the center position. 
Connect the RTP to the Reacti-Therm Module (left side). When the Heat knob is turned on (step 4), the Temp. Probe 
light will illuminate to indicate that the RTP is connected and operating as the thermal sensor. 

Note:

 The Heat knob must be in the off position when connecting or disconnecting (plugging or unplugging) the RTP. 

3.

 

Place the RTP tip into the aluminum block thermometer well and insert a thermometer (200°C range) into a reaction vial 
placed in one of the block wells. For in-vial placement of the RTP, fill the vial sufficiently with mineral oil or other 
suitable liquid so that the probe tip is immersed at least 2.5 cm to ensure adequate thermal contact with the vial.  

4.

 

Turn the Heat knob clockwise to set the desired target block temperature (°C), as indicated in the LED display above the 
Heat knob. The displayed value will continue to flash (blink) until the block temperature approaches the set temperature.  

Example:

 To set temperature to 100°C, turn Heat knob until the value “100” appears (flashing) in the LED display. 

5.

 

Allow 15-30 minutes for the block to approach the set temperature, at which point the LED display stops flashing the set 
temperature and begins reporting (without flashing) the block temperature recorded by the RTP. The actual time required 
for the unit to reach a given temperature may vary depending on the block used and the ambient (room) temperature. 

6.

 

When the LED display stops flashing, monitor the displayed temperature for an additional 15 minutes to ensure that the 
it has fully stabilized (i.e., that the in-vial thermometer reading remains unchanged for several minutes). Note that the 
thermometer reading, although stabilized, might still be different from the set temperature reported by the RTP. 

7.

 

Once the block has stabilized at a given temperature, turn the Display Adjust knob until the LED display value matches 
the block temperature indicated by the in-vial thermometer. 

Example:

 Suppose the Heat knob had been set to 100°C (step 4) as indicated by the LED display. After 30 minutes, the 

block temperature measures 98°C (as indicated by the in-vial thermometer) and has remained unchanged for several 
minutes. Harmonize the reported (displayed RTP) value to the actual (thermometer) temperature by turning the Display 
Adjust knob until the LED display value equals 98°C. 

8.

 

If the Display Adjust was used in step 7, re-position the Heat knob to the original target temperature. 

Example:

 Having changed the Display Adjust knob until the LED display value equals 98°C, now turn the Heat knob so 

that the set temperature (LED display) equals 100°C. 

9.

 

After several minutes, the block will stabilize to the desired temperature. 

Note:

 Once calibration has been performed for a given aluminum block, vial size and ambient condition, setting the 

Heat knob to the same set temperature with the Display Adjust knob at the same dial position for subsequent uses 
will reproduce the same heating performance. You may wish to mark or note the Display Adjust dial position for 
future reference when using this same heating application.  

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание Reacti-Therm I

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...240 V AC Input Voltage Tolerance 10 10 Wattage Maximum 130 W 260 W Frequency 50 60 Hz 50 60 Hz Fuse Type 3 A 250 V Fast acting 3 A 250 V Fast acting Size 5 20 mm 5 20 mm Location Inside cord inlet hou...

Страница 3: ...intenance of electrical apparatus for use in potentially explosive atmospheres When heating certain substances the liberation of hazardous gases may require the use of a fume hood or other suitable ex...

Страница 4: ...counter clockwise Set the Temperature Display Adjustment knob Display Adjust knob to the center position 3 Place a thermometer 200 C range Product No 18915 into the aluminum block thermometer well 4...

Страница 5: ...rmometer 200 C range into a reaction vial placed in one of the block wells For in vial placement of the RTP fill the vial sufficiently with mineral oil or other suitable liquid so that the probe tip i...

Страница 6: ...Hydrolysis Tubes 5 ml 29564 Vacuum Hydrolysis Tubes 20 ml This product Product is warranted to operate or perform substantially in conformance with published Product specifications in effect at the t...

Страница 7: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Отзывы: