![Thermo Scientific RadEye GF Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/thermo-scientific/radeye-gf/radeye-gf_operating-instructions-manual_1105259078.webp)
7-6
DB-083 E
RadEye GF-Series
Le/Ff 20.07.2015
7.7
RadEye’s battery status is lower than expected
Please check if the correct battery type is selected under “set-
tings” / “Batt. type” and change the type in case. Rechargeable
batteries supply a voltage of 1.2V each, while Alkaline batter-
ies provide 1.5V.
Rechargeable batteries offer also a more steep drop from ac-
ceptable battery voltage to “battery low” than Alkaline types.
7.8
History data shows the wrong time and date
See chapter 7-5.