5-2
DB-083 E
RadEye GF-Series
Le/Ff 20.07.2015
Error high voltage generation.
No detector pulse within the last 30
minutes.
EEPROM with calibration data shows
EEPROM Read or EEPROM Write
error.
A Watchdog Error indicates that the
microcontroller has problems to work
on its tasks in a given timeframe.
Reasons are strong electromagnetic
pulses, firmware errors or hardware
issues.
The BTcom cover failed to initialize
the Bluetooth module.