![Thermo Scientific RadEye GF Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/thermo-scientific/radeye-gf/radeye-gf_operating-instructions-manual_1105259050.webp)
5-4
DB-083 E
RadEye GF-Series
Le/Ff 20.07.2015
5.4
Low Battery
If the battery voltage is below the configured threshold (see
chapter 4.12) the following warning appears:
The beeper generates a single pulse every 32s. This acoustic
warning can be suppressed by pressing the alarm acknowledge
key. After 8h this warning comes up again. The battery needs
to be changed. However, the RadEye can still be operated for
several hours.
If the battery voltage falls below 2V, the battery symbol starts
flashing. An acoustic warning is generated every 8s and can’t
be acknowledged. The battery needs to be changed as soon as
possible.
Содержание RadEye GF
Страница 1: ...Operating Instructions DB 083 E RadEye GF Series GF GF 10 Gamma Survey Meters...
Страница 2: ......
Страница 14: ......
Страница 52: ......
Страница 71: ...RadEye GF Series DB 083 E 6 19 Le Ff 20 07 2015 Figure 6 14 Logbook...
Страница 72: ......
Страница 82: ......
Страница 101: ...RadEye GF Series DB 083 E 11 7 Le Ff 20 07 2015 Diagram 11 3 Angular response horizontal plane RadEye GF...
Страница 102: ...11 8 DB 083 E RadEye GF Series Le Ff 20 07 2015 Diagram 11 4 Angular response horizontal plane RadEye GF 10...
Страница 103: ...RadEye GF Series DB 083 E 11 9 Le Ff 20 07 2015 Diagram 11 5 Angular response vertical plane RadEye GF...
Страница 104: ...11 10 DB 083 E RadEye GF Series Le Ff 20 07 2015 Diagram 11 6 Angular response vertical plane RadEye GF 10...
Страница 108: ......
Страница 112: ......
Страница 113: ......