Thermo Scientific PowerEase PS0300 Скачать руководство пользователя страница 9

PowerEase

®

 300W Power Supply User Guide

  

 

 9

 

 

36B

Constant 

operation protocol 

Instructions for operating the PowerEase

®

 300W Power Supply in the 

Constant Mode 

are 

provided in this section.  The 

Constant Voltage, Current, or Power Mode 

allows you to specify 

a voltage limit, current limit and power limit to be used during the entire duration of 
electrophoresis.  Review the guidelines provided in this manual before starting electrophoresis 
using the PowerEase

®

 300W Power Supply. 

A basic 

Constant Voltage, Current Mode, or Constant Power 

operating procedure of the 

PowerEase

®

 300W Power Supply is provided below. We recommend reading the guidelines 

provided in this manual for best results before starting an operation. 

Note:

 The PowerEase

®

 300W is a high throughput power supply.  For optimal performance, Life 

Technologies recommends using Constant Voltage when running 1–2 protein gels or blots. 

1.

 

Use the power switch on the rear side of the instrument to turn on the PowerEase

®

 300W 

Power Supply. On display screen

 Constant Mode 

and

 Program Mode

 will appear. 

2.

 

With the 

Constant Mode

 highlighted press the 

Mode

 key.  

3.

 

The 

Mode

 key toggles editing between 

VOLT

TIME

 (hours/minutes), 

AMP

, and 

WAT

  

4.

 

Press the 

Constant

 key to select the operation modes: Constant Voltage, Constant Current, or 

Constant Power from the display screen. 

 

The screen displays the selected operation mode on the left side of the display. 

5.

 

Use the 

 

 arrow keys to set the selected operation [voltage (VOLT), current (AMP), or 

Power (WAT)] to the appropriate values. 

6.

 

Use the 

Mode

 key to select TIME and use 

 

 arrow key to set the time (hours/minutes) to 

specify the duration of the electrophoresis.  

7.

 

Press 

Start/Pause 

key to start electrophoresis.  The Red LED illuminates when the 

electrophoresis is running. 

8.

 

Press the 

Start/Pause

 key again to temporarily interrupt power to ongoing electrophoresis 

without terminating the operation along with LED flashing. 

9.

 

Press 

Start/Pause 

key to restart electrophoresis. 

10.

 

Press 

Stop 

key to stop the electrophoresis. 

Note:

 If the parameters need to be changed during the run press 

Start/Pause

, enter the changes, and then 

press 

Start/Pause

 once again to restart the operation. 

Note

: After stopping and restarting an operation, the timer resets to the selected time.  

 

37B

Basic 

programming 
protocol 

Instructions for operating the PowerEase

®

 300W Power Supply in the 

Program Mode 

are 

provided in this section.  The 

Program Mode 

allows user to vary levels in voltage (V), current 

(mA), and power (W) during specified periods of time as discrete changes (STEP) for up to 10 
steps, depending upon electrophoresis needs. 

A basic 

Program Mode 

operating procedure of the PowerEase

®

 300W Power Supply is provided 

below. For best result, read the guidelines provided in this manual before starting an operation. 

 

 

 

Содержание PowerEase PS0300

Страница 1: ...PowerEase 300W Power Supply Catalog Numbers PS0300 PS0301 PS0302 Publication Number MAN0008988 Revision D 0 Research Use Only Not for use in diagnostic procedures...

Страница 2: ...SE OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR NON INFRINGEMENT TO THE EXTENT ALLOWED BY LAW IN NO EVENT SHALL LIFE TECHNOLOGIES AND OR ITS AFFILIATE S BE LIABLE WHETHER IN CONTRACT TORT WA...

Страница 3: ...peration modes 7 15BLimiting parameter settings 11 2BAppendix A 12 16BGeneral safety 12 Instrument safety 12 17BExplanation of symbols and warnings 13 1Product regulatory compliance 13 3BAppendix B 15...

Страница 4: ...ims must be filed with the carrier keep the supplied box for inspection The warranty does not cover in transit damage To ensure safe reliable operation always operate the PowerEase 300W Power Supply i...

Страница 5: ...wer Supply Constant Voltage Power or Current Capable of running multiple electrophoresis units LCD display with clear menu prompts for easy use Programming capabilities for limiting voltage V current...

Страница 6: ...e the cursor position down and decrease values when settings are flashing Mode Select Constant Mode or Program Mode The Mode key cycles between voltage current power and time settings allowing values...

Страница 7: ...riations in buffer conditions can result in exceeding the power supply s maximum voltage current or power output capacity and produce undesirable variations in performance 33BRecommendation The durati...

Страница 8: ...Mode the unit defaults to the Constant Voltage mode and Volts V displays on the left side of the screen in large characters The controlling value Volts V Amps A or Watts W is always displayed on the...

Страница 9: ...Power from the display screen The screen displays the selected operation mode on the left side of the display 5 Use the arrow keys to set the selected operation voltage VOLT current AMP or Power WAT t...

Страница 10: ...of the run press the Mode key 7 Voltage will appear on the display to enter voltage press Mode key 8 To change voltage use the arrow keys 9 Select AMP or WAT by pressing the Mode key until desired par...

Страница 11: ...ng margin of safety over time The same voltage setting can be used regardless of the number or thickness of gels being electrophoresed 42BCurrent limiting Discontinuous buffer systems and to a lesser...

Страница 12: ...indicator light at the Start Stop key is on or when RUNNING is displayed on the screen 2 WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus 3 ALWAYS make...

Страница 13: ...tri s lectif CAUTION Pour minimiser les consequences n gatives sur l environnement la suite de l limination de d chets lectroniques ne pas liminer ce d chet lectronique avec les d chets usuels non so...

Страница 14: ...uency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely...

Страница 15: ...cell Correct the condition by making sure the tape is removed from the pre cast gel buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation s...

Страница 16: ...oved during a run Buffer leaking in a connected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Current set too low Change in constant mode Vo...

Страница 17: ...ghly conductive chloride ions in the gel are gradually replaced by the slower moving less conductive glycine ions from the running buffer as the gel runs As a result the resistance of the gel increase...

Страница 18: ...ant Current Constant Power 1V step 10 mA step 1 watt step Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause function Yes Safety features No Load Detec...

Страница 19: ...Appendix F To replace the fuse 1 Turn off the main power switch at the rear of the PowerEase 300W Power Supply and detach the power cord from the rear of the PowerEase 300W Power Supply 2 Open the fu...

Страница 20: ...tomer training Download software updates and patches 57BSafety Data Sheets SDS Safety Data Sheets SDSs are available at HUwww lifetechnologies com supportUH 58BCertificate of Analysis The Certificate...

Отзывы: