Thermo Scientific PowerEase PS0300 Скачать руководство пользователя страница 15

PowerEase

®

 300W Power Supply User Guide

  

 

15

 

 

3B

Appendix B 

19B

Troubleshooting 

Observation 

Possible cause 

Recommended action 

The LCD screen 
remains blank and the 
fan does not run when 
the power is turned on 

AC power cord is not connected 

Check AC power cord connections at both ends.  

Use the correct cord. 

The fuse has blown 

Replace the fuse 

Operation stops with 
alarm: The screen 
displays “

NO LOAD

” 

Electrophoresis leads are not 
connected to the power supply or 
to the electrophoresis unit(s), or 
there is a broken circuit in the 
electrophoresis cell 

Check the connections to the power supply and on 
your electrophoresis cell to make sure the 
connection is intact; check condition of wires in 
electrophoresis unit. Close the circuit by 
reconnecting the cables. Press 

Start/Pause

 to 

restart the run. 

High resistance due to tape left on 
a pre-cast gel, incorrect buffer 
concentration, or incorrect buffer 
volumes in the electrophoresis cell 

Correct the condition by making sure the tape is 
removed from the pre-cast gel, buffers are 
prepared correctly, and the recommended volume 
of buffer is added to the electrophoresis unit. 

Operation stops with 
alarm: Display shows 

OVER VOLTAGE

” 

Circuit is interrupted 

 

Verify that the running buffer is correct. 

 

Verify the all cables are attached correctly 

 

Turn the Power switch off and on again; 
restart application. 

 

If you cannot restart the instrument, turn off 
the power, disconnect the power cord from the 
outlet, and contact Technical Service. 

Operation stops with 
alarm: Display shows 

LEAKAGE

” 

Ground leak detected during run 

Check the electrophoresis system for improper 
grounding. Restart the power supply by turning the 
Power switch off and on. 

Operation stops with 
alarm: 

Display shows “

OVER 

TEMP

” 

Power supply is overheating 

 

Turn off power supply. Check for sufficient 
airflow around the power supply fan. After 
cooling down, restart the power supply by 
turning the Power switch to the on position. 

 

If you cannot restart the instrument, turn off 
the power, disconnect the power cord from the 
outlet, and contact Technical Service. 

 

 

 

Содержание PowerEase PS0300

Страница 1: ...PowerEase 300W Power Supply Catalog Numbers PS0300 PS0301 PS0302 Publication Number MAN0008988 Revision D 0 Research Use Only Not for use in diagnostic procedures...

Страница 2: ...SE OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR NON INFRINGEMENT TO THE EXTENT ALLOWED BY LAW IN NO EVENT SHALL LIFE TECHNOLOGIES AND OR ITS AFFILIATE S BE LIABLE WHETHER IN CONTRACT TORT WA...

Страница 3: ...peration modes 7 15BLimiting parameter settings 11 2BAppendix A 12 16BGeneral safety 12 Instrument safety 12 17BExplanation of symbols and warnings 13 1Product regulatory compliance 13 3BAppendix B 15...

Страница 4: ...ims must be filed with the carrier keep the supplied box for inspection The warranty does not cover in transit damage To ensure safe reliable operation always operate the PowerEase 300W Power Supply i...

Страница 5: ...wer Supply Constant Voltage Power or Current Capable of running multiple electrophoresis units LCD display with clear menu prompts for easy use Programming capabilities for limiting voltage V current...

Страница 6: ...e the cursor position down and decrease values when settings are flashing Mode Select Constant Mode or Program Mode The Mode key cycles between voltage current power and time settings allowing values...

Страница 7: ...riations in buffer conditions can result in exceeding the power supply s maximum voltage current or power output capacity and produce undesirable variations in performance 33BRecommendation The durati...

Страница 8: ...Mode the unit defaults to the Constant Voltage mode and Volts V displays on the left side of the screen in large characters The controlling value Volts V Amps A or Watts W is always displayed on the...

Страница 9: ...Power from the display screen The screen displays the selected operation mode on the left side of the display 5 Use the arrow keys to set the selected operation voltage VOLT current AMP or Power WAT t...

Страница 10: ...of the run press the Mode key 7 Voltage will appear on the display to enter voltage press Mode key 8 To change voltage use the arrow keys 9 Select AMP or WAT by pressing the Mode key until desired par...

Страница 11: ...ng margin of safety over time The same voltage setting can be used regardless of the number or thickness of gels being electrophoresed 42BCurrent limiting Discontinuous buffer systems and to a lesser...

Страница 12: ...indicator light at the Start Stop key is on or when RUNNING is displayed on the screen 2 WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus 3 ALWAYS make...

Страница 13: ...tri s lectif CAUTION Pour minimiser les consequences n gatives sur l environnement la suite de l limination de d chets lectroniques ne pas liminer ce d chet lectronique avec les d chets usuels non so...

Страница 14: ...uency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely...

Страница 15: ...cell Correct the condition by making sure the tape is removed from the pre cast gel buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation s...

Страница 16: ...oved during a run Buffer leaking in a connected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Current set too low Change in constant mode Vo...

Страница 17: ...ghly conductive chloride ions in the gel are gradually replaced by the slower moving less conductive glycine ions from the running buffer as the gel runs As a result the resistance of the gel increase...

Страница 18: ...ant Current Constant Power 1V step 10 mA step 1 watt step Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause function Yes Safety features No Load Detec...

Страница 19: ...Appendix F To replace the fuse 1 Turn off the main power switch at the rear of the PowerEase 300W Power Supply and detach the power cord from the rear of the PowerEase 300W Power Supply 2 Open the fu...

Страница 20: ...tomer training Download software updates and patches 57BSafety Data Sheets SDS Safety Data Sheets SDSs are available at HUwww lifetechnologies com supportUH 58BCertificate of Analysis The Certificate...