PowerEase
®
300W Power Supply User Guide
2B
Appendix A
16B
General safety
44B
WARNING!
Protection impairment if used in a manner not specified by the manufacturer.
Instrument safety
44B
Avoid electrical
shock
The PowerEase
®
300W Power Supply produces up to 3A or 250 voltage outputs which are
electrically isolated from ground to reduce the risk of electrical shock to the user. Follow the
guidelines below to ensure safe operation of the unit.
The PowerEase
®
300W Power Supply has been designed for use with electrophoresis cells with
shielded banana plugs thus minimizing any potential shock hazard to the user. Life Technologies
recommends against the use of unshielded banana plugs.
To avoid electrical shock:
1.
NEVER connect or disconnect wire leads from the power jacks when the red indicator light at
the Start/Stop key is on or when “RUNNING” is displayed on the screen.
2.
WAIT at least 5 seconds after stopping a run before handling output leads or connected
apparatus.
3.
ALWAYS make sure that hands, work area, and instruments are
clean
and
dry
before making
any connections or operating the power supply.
4.
ALWAYS connect the power supply to a properly grounded AC outlet.
45B
Avoid damage to
the instrument
1.
For proper ventilation, leave at least 10 cm of space behind the instrument, and at least 5 cm of
space on each side.
2.
Do not operate the power supply in high humidity environments (> 95%), or where
condensation may occur.
3.
To avoid condensation after operating the power supply in a cold room, place the unit in a
sealed plastic bag prior to removing the power supply from the cold room. After removing the
power supply from the cold room, allow at least 2 hours for the unit to equilibrate to room
temperature before removing the plastic bag and operating the unit.