background image

Thermo Scientific

ROSS Ultra, ROSS Ultra Triode and ROSS pH Electrodes 

|

 

9

Thermo Scientific Orion

and water saturated with KCl. The ratio of methanol to water should be 

similar to the sample composition. For tips on measuring pH in non-

aqueous samples, see Thermo Scientific Water Analysis Instruments 

Application Note 007.

If using the 8220BNWP ROSS micro pH electrode, great care must 

be taken to thoroughly rinse out the reference chamber and junction 

before changing the reference solutions.

Measuring Hints 

• 

Always use fresh buffers for calibration. Choose buffers that are one 

to four pH units apart. 

• 

Check the electrode slope daily by performing a two buffer 

calibration. The slope should be 92 to 102%. 

• 

Except as noted in the 

Sample Requirements

 section, only use 

ROSS fill solution, Cat. No. 810007. Do not use any fill solution that 

may contain silver. 

• 

If a refillable electrode is used, always remove the fill hole cover 

during measurements to ensure a uniform flow of fill solution. 

• 

If a ROSS Ultra Triode is used, immerse the electrode at least 2 

inches (5 cm) in solution for best temperature response.

• 

Between measurements, rinse the electrode with distilled water and 

then with the next solution to be measured. 

• 

Stir all buffers and samples at a uniform rate. 

• 

Place a piece of insulating material, such as polystyrene foam or 

cardboard, between the magnetic stir plate and beaker to prevent 

measurement errors from the transfer of heat to the sample. Since 

ROSS electrodes respond quickly to temperature changes, pH 

changes due to temperature variations will be noticed. 

• 

To reduce the chance of error due to polarization, avoid rubbing or 

wiping the sensing bulb. Use a lint-free tissue and gently blot the 

sensing bulb.

• 

If a refillable electrode is used in dirty or viscous samples or 

the electrode response becomes sluggish, empty the electrode 

completely and hold the junction under running water. Empty any 

water from the electrode and refill it with fresh ROSS fill solution. 

For a thorough cleaning method, refer to the 

Electrode Cleaning 

Procedures

 section.

• 

If a Sure-Flow electrode is used in dirty or viscous samples or 

the electrode response becomes sluggish, empty the electrode 

completely and hold the junction open under running water. Empty 

any water from the electrode and refill it with fresh ROSS fill 

solution. Press down on the electrode cap to let a few drops of the 

ROSS fill solution flow out of the electrode and then replenish any 

lost solution. For a thorough cleaning method, refer to the 

Electrode 

Cleaning Procedures

 section.

• 

Handle the 8220BNWP ROSS micro pH electrode with care. Do 

not touch the pH bulb and stem against the bottom or walls of the 

sample containers. 

Содержание Orion ROSS

Страница 1: ...ROSS Ultra ROSS Ultra Triode and ROSS pH Electrodes 263745 001 Revision B November 2014 Thermo Scientific Orion U ser Manual...

Страница 2: ......

Страница 3: ...ode Calibration 10 General Calibration Procedure 10 One Buffer Calibration 10 Two Buffer Calibration 11 pH Measurements 12 Electrode Storage 12 Electrode Maintenance 12 Electrode Cleaning Procedures 1...

Страница 4: ...s are three to five times more precise than those obtained with conventional electrodes Over the lifetime of the electrode ROSS reference potential drifts less than an average 0 12mV per day which cor...

Страница 5: ...ientific Orion Star A200 series Thermo Scientific Orion Star A100 series or Thermo Scientific Orion Dual Star meter with pH measurement capability such as the Versa Star benchtop meter with pH module...

Страница 6: ...3 The electrode may arrive with the chamber partially filled with electrolyte solution Please fill the chamber with electrolyte solution before use Uncover the fill hole by removing the tape and add...

Страница 7: ...an adequate flow rate the level of fill solution must always be above the reference junction and at least one inch above the sample level The fill hole should remain open whenever the electrode is in...

Страница 8: ...on aqueous solutions and solutions that contain organic solvents A minimum of 20 water must be present in the sample for the best measurement results If there is a great deal of drift when using the R...

Страница 9: ...a uniform rate Place a piece of insulating material such as polystyrene foam or cardboard between the magnetic stir plate and beaker to prevent measurement errors from the transfer of heat to the sam...

Страница 10: ...the meter is set to automatically recognize the pH buffer at the measured temperature and an ATC probe is used you do not need to manually adjust the pH values of the buffers One Buffer Calibration 1...

Страница 11: ...rode into the first buffer When the reading is stable set the meter to the pH value of the buffer at the measured temperature Refer to the meter user guide for a detailed procedure Table 1 provides pH...

Страница 12: ...flush it with fresh ROSS fill solution and refill the chamber with fresh ROSS fill solution Electrode Cleaning Procedures General Cleaning We offer a variety of Orion pH cleaning solutions for your c...

Страница 13: ...n Cat No 910001 for at least one hour Cleaning Solutions Cat No 900021 WA pH cleaning solution A for removing protein contaminants Cat No 900022 WA pH cleaning solution B for removing bacterial contam...

Страница 14: ...C probe are placed into a sample that varies significantly in temperature the readings can drift for two reasons First the temperature response of the electrode and temperature probe may not be simila...

Страница 15: ...m precision a three point temperature calibration is recommended Do not perform a one point temperature calibration if measured solutions will be below 20 C The following tables illustrate the expecte...

Страница 16: ...s and at higher temperatures Because the pH membrane is composed of special low sodium error glass error due to sodium is negligible when measuring at pH values less than 12 When measuring at pH value...

Страница 17: ...lectrode into fresh pH 4 buffer When the reading is stable record the mV value of the pH 4 buffer The mV value should be in the range of 150 to 210 mV 4 Calculate the absolute mV difference between th...

Страница 18: ...Mercury Hg Cadmium Cd Sensor body with cable assembly X 0 0 Plastic hardware 0 0 0 Part Name Toxic or Hazardous Substances and Elements Hexavalent Chromium Cr VI Polybrominated biphenyls PBB Polybromi...

Страница 19: ...combination pH electrode with epoxy body and semi micro tip waterproof BNC connector 8135BNUWP ROSS Ultra combination pH electrode with epoxy body and flat surface tip waterproof BNC connector 8156BN...

Страница 20: ...ctrode with epoxy body waterproof BNC connector 8165DN ROSS Sure Flow combination pH electrode with epoxy body E DIN waterproof connector 8172BNWP ROSS Sure Flow combination pH electrode with glass bo...

Страница 21: ...scientific com water for additional accessories buffers and buffer sizes Specifications pH Range 0 to 14 pH pH Precision 0 01 pH Temperature Range 0 to 100 C ROSS Ultra Triode Temperature Accuracy 1 C...

Страница 22: ...8102BN 8104BN 8135BN 815600 8162SC 8165BNWP 8172BNWP 8302BNURCA 8107BNURCA 8157BNURCA 120 mm length 12 mm diameter 8103BNUWP 8103BN 8103SC 165 mm length 95 mm tip length 6 mm tip diameter 8115BNUWP 8...

Страница 23: ......

Страница 24: ...moscientific com water Netherlands Tel 31 020 4936270 info water uk thermo com China Tel 86 21 68654588 wai asia thermofisher com India Tel 91 22 4157 8800 wai asia thermofisher com Singapore Tel 65 6...

Отзывы: