Orion 8010cX Ammonia Analyzer
Common Faults and Troubleshooting Measures | 45
5.After all the above steps have been checked and the problem is solved,
rerun the analysis procedure to see if the fault has been eliminated. If
the fault still occurs, contact the after-sales personnel to confirm the
fault and repair it.
20
Lack of standard sample (E8)
1.Check whether the standard sample in the standard sample bottle is
used up or insufficient;
2.Whether the standard sample injection tube has been inserted into the
bottom of the standard sample bottle;
3.Check whether the solution back flow occurs at the place where the
liquid inlet tube of standard sample is connected to the port 2 of the
electronic rotary valve; if yes, the connection terminal at this place need
to be tightened;
4.After all the above steps have been checked and the problem is solved,
rerun the analysis procedure to see if the fault has been eliminated. If
the fault still occurs, contact the after-sales personnel to confirm the
fault and repair it.
21
Lack of cleaning solution (E11)
1.Check whether the cleaning solution in the cleaning solution bottle is
used up or insufficient;
2.Whether the cleaning solution tube has been inserted into the bottom of
the cleaning solution bottle;
3.Check whether the solution back flow occurs at the place where the
liquid inlet tube of cleaning solution is connected to the port 1 of the
electronic rotary valve; if yes, the connection terminal at this place need
to be tightened;
4.After all the above steps have been checked and the problem is solved,
rerun the analysis procedure to see if the fault has been eliminated. If
the fault still occurs, contact the after-sales personnel to confirm the
fault and repair it.
Table 12. Alarm/Fault Information and Troubleshooting Methods
No.
Alarm Information
Troubleshooting Methods