Orion 8010cX Ammonia Analyzer
Common Faults and Troubleshooting Measures | 41
4
Unstable temperature of the
reaction cell (W1)
1.When the instrument is not working, observe the temperature of the
reaction cell. The temperature change of the reaction cell should be
within ±1°C;
2.If the temperature of the reaction cell changes greatly, the instrument
may be failure. Contact the after-sales personnel to confirm the fault
and repair it.
5
High/low ambient temperature
(W2, W3)
1.Observe the ambient temperature, its changes should be within the
range of 5°C ~ 45°C; if the temperature change exceeds the above
range, turn on the air conditioner and other equipment to adjust the
indoor temperature to the set range, otherwise the working
performance of the instrument will be affected and even the hardware
of the instrument will be damaged;
2.If the ambient temperature changes greatly, the instrument may be
failure. Contact the after-sales personnel to confirm the fault and repair
it.
6
High/low temperature of
reaction cell (E15, E16)
1.When the instrument is not working, observe the temperature of the
reaction cell. The temperature of the reaction cell should be within the
range of the set temperature ±5°C;
2.If the temperature of the reaction cell deviates too much from the set
temperature, the instrument may be failure, contact the after-sales
personnel to confirm the fault and repair it.
7
Expiration of service life of
injection pump three-way valve
(W4)
1.The service life of the injection pump three-way valve is designed to be
1 year. Extended service will lead to instability of the system, abnormal
test results and even damage to the injection pump;
2.if the service life of the injection pump three-way valve expires, contact
the after-sales personnel for maintenance and replacement.
8
Expiration of service life of
syringe pump (W4)
1.The service life of the syringe pump is designed to be half a year.
Extended service will lead to instability of the system, abnormal test
results and even damage to the injection pump;
2.If the service life of the syringe pump expires, contact the after-sales
personnel for maintenance and replacement.
9
Expiration of service life of rotary
valve (W7)
1.The service life of the multi-way rotary valve is designed to be three
years. Extended service will cause instability of the system, abnormal
test results, and even damage to the instrument;
2.If the service life of the rotary valve expires, contact the after-sales
personnel for maintenance and replacement.
Table 12. Alarm/Fault Information and Troubleshooting Methods
No.
Alarm Information
Troubleshooting Methods