24 | System Startup
Orion 8010cX Ammonia Analyzer
Installation of Reagent Bottle and
Standard Solution Bottle
For the installation of reagent bottle and standard solution
bottle, it is recommended to operate in the mode shown in
Figure 21. Recommended installation method of
Reagent Bottle
Installation of Tested Water Sample
After the tested water sample tube is led out from the rubber
stopper at the lower left side of the instrument, it is inserted
into the top inlet of the flow cell. Refer to
In the flow cell (or similar equipment), the water sample
needs to be retained. It is recommended that the water
sample shall be fully discharged before the end of
measurement to ensure that the equipment needs
special operations such as range switching.
Installation of Liquid Waste Tank
and Deionized Water Tank
In total, the instrument has two liquid waste tubes for
discharge. One is connected to the chemical liquid waste
tank and the other is connected to the cleaning liquid waste
tank. It is also required that the liquid waste tube can be
correctly connected to the lid of the liquid waste tank to
prevent liquid waste from flowing out and causing safety
problems.
Both the deionized water inlet and two liquid waste discharge
ports are at the bottom of the instrument, so both the
deionized water tank and liquid waste tank need to be placed
below the instrument.
Main steps of liquid waste tubes installation:
a.Two liquid waste tubes pass through two rubber sealing
covers at the bottom of the instrument cabinet;
b.Adjust and cut the liquid waste tube outside the instrument
to a suitable length (according to the specific position of the
liquid waste tank), fix it on the lid of the liquid waste tank
and extend it into the liquid waste tank.
Note:
• Only trained and qualified personnel should perform liquid
waste related operations;
• The user shall regularly treat the chemical liquid waste tank
as required and add deionized water as required (it is
recommended that the maximum interval should not
exceed one month).
• The liquid waste tubes should not be immersed below the
level of liquid waste to avoid the suck-back of liquid wastes.