Terms Definition
Section 8
2230XP Silica Analyzer
Thermo Scientific Orion
63
Terms
Description
Possible Values, Options
Stir Time
The amount of time the fluid is
stirred after pumping the reagent.
0 to 999 seconds
DATE
Format
Selectable date format.
MM/DD/YY, YY/MM/DD, DD/MM/YY
Current Date
Present date, validated depending
on the format.
Legible date
TIME
Current Time
Present time.
Legible time depending on time format
Format
Selectable time format.
12 hr, 24 hr
RECOVERY AFTER POWER FAILURE
Power Fail Duration
Recovery duration after power fail.
0 to 99.99 hr
Power Fail Clean Cycle
Performing clean cycle after
recovering from power fail.
Enable, Disable
ETHERNET PARAMETERS
IP Address
Instrument IP address.
Editable
Server IP
Network ID
RS-232 PARAMETERS
Baud Rate
The rate at which the data is
transferred through serial port.
115200 (default)
LANGUAGE
Language
Language in the instrument.
English, other languages, consult
factory
REVIEW / RESET ALARMS
Reset All Alarms?
Resets all the raised alarms,
which are shown below this option.
Reset All Alarms?
Reset All Faults?
Resets all the faults, which are
shown below this option.
None
REVIEW CALIBRATION RESULTS
Display
Displays the calibration results
for the selected data range.
None
Prints the Calibration results on to
the PC for the selected data range.
None
Data Range
The time interval in which the
calibration results are to be
retrieved.
Last 12 hr, 1 Day, 2 Days, 3 Days,
1 Week, 2 Weeks, 3 weeks and All
REVIEW ANALYSIS RESULTS
Display
Displays the analysis results
for the selected data range.
None
View Graphs
Displays the Linear graph for the
selected data range of Analysis
records.
None
Prints the analysis results on to
the PC for the selected data range.
None
Data Range
The time interval in which the
analysis results are to be retrieved.
Last 12 hr, 1 Day, 2 Days, 3 Days,
1 Week, 2 Weeks, 3 weeks and All