Section 3
Installation
Environmental Considerations
18
Thermo Scientific Orion
2230XP Silica Analyzer
SAMPLE IN
SAMPLE RETURN
FLOW DIRECTION
FILTERED AND OIL FREE
AIR PURGE SUPPLY LINE
CONNECT TO
ATMOSPHERIC DRAIN
FLOW DIRECTION
SHUT-OFF
VALVE
AC MAINS
1
0
POWER CIRCUIT BREAKER
MUST BE CLEARLY IDENTIFIED
INSTALL HEIGHT TO
OPERATOR EYE LEVEL
PROCESS LINE
USER TO INSTALL A 40 to 60 MICRON
FILTER, SIZE IS BASED ON SAMPLE
CONDITION, SET FLOW RATE
BETWEEN : 50 TO 500 ML/MIN
AC POWER
SAMPLE CHAMBER
SAMPLE (BACK) N.C.
STANDARD (FRONT) N.O.
RGT1
RGT3
STD
RGT2
RGT3
DRAIN
SAMPLE (INLET)
RGT1
Figure 2: Mounting dimensions
SAMPLE CHAMBER
SAMPLE IN 1/4 ''
SAMPLE RETURN 1/4 ''
DRAIN 3/8 ''
VENT LINES
AIR PURGE
AIR PURGE
AIR VENT
I/O SIGNALS
POWER IN
ALLOW 20 INCHES
(510 mm) CLEARANCE
FOR DOOR SWING
70mm(2.75 ")
304.8mm (12.0")
67
4.6
mm (26
.5
")
132mm (5.19 ")
25mm (0.98 ")
30mm (1.18 ")
33
0.3m
m
(1
3.0
")
41mm (1.62 ")
162mm (6.4")
69
8m
m
(27
.5
")
15
2m
m
(6
")
406mm (16")
SAMPLE
STANDARD