Troubleshooting
VI-7
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
• Press
to start display test.
1. All the display segments will light up and the system will cycle through
the LED colors.
2. Verify that all the display segments are on and that the LED colors are
functional and press
.
3. All the display segments will turn off except for the marquee display,
which should show
PRESS ENTER TO CONTINUE
.
4. Verify that all the display segments are off and press
.
5. The system will show a counting number pattern on both lines of the
main display.
6. Verify that the display segments are functional and press
to
proceed to the keypad test.
Display Test
Default
Operator Action
Scrolling Marquee
Main Display
Notes
Display Test
PRESS ENTER FOR THE DISPLAY TEST
dISP
tESt
Press
to start the test
Verify that all of the display segments are
on and that the LED colors are functional
All segments on
All segments on
System will
cycle through
the LED colors
Press
to proceed to the next part
of the test
Verify that all of the display segments
are off
PRESS ENTER TO CONTINUE
All segments off
Press
to proceed to the next part
of the test
Verify that all of the numeric display
segments are functional
PRESS ENTER TO CONTINUE
Count pattern
Count pattern
Press
to proceed to the next menu