Analyzer Preparation
II-11
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
1. Unpack the electrode cables.
2. Feed the tinned wires through the cable gland assemblies with the holes
(2 or 1).
3. Follow the terminal assignments shown in
Figure II-6
for the proper
electrode cable wiring location.
1. Open the ferrite using a flat tip screwdriver to lift the latch of
the ferrite.
2. Feed the cable wires through the center of the ferrite and then loop
the cable wires around the ferrite core and through the center of the
ferrite again.
3. Place the ferrite at the bottom of the analyzer chassis, near its edge.
Adjust the ferrite location on the cable so the ferrite is near the top of
the cable.
Installation of New
Electrode Cables
Ferrite Installation
Figure II-8
Ferrite Installation