![Thermo Scientific NIVETEC Orion 2109XP Скачать руководство пользователя страница 116](http://html.mh-extra.com/html/thermo-scientific/nivetec-orion-2109xp/nivetec-orion-2109xp_user-manual_1105205116.webp)
A-5
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
Appendix
Outputs
Number of Analog Outputs
Two, one dedicated to calcium, one dedicated to
temperature and shared ground
Analog Outputs
Galvanically isolated
Output Selections
0 to 20 mA or 4 to 20 mA
Programmable Range
Yes
Channel Assignments
Calcium and temperature
Relative Accuracy
± (0.05 mA + 0.5%)
Maximum Load
500 ohm or 10 V
Log And Linear Output Options
Yes, operator selectable
Alarm Outputs
Number of Relay Outputs
3
Maximum Relay Load
250 VAC/5 A, 30 VDC/5 A, resistive load only
Minimum Value Alarm
Yes
Maximum Value Alarm
Yes
Error Alarm
Yes
Calibration/offline Alarm
Yes
Programmable Min. and Max. Values
Yes
Analyzer Features
Non-volatile Memory
Yes
Battery Backup (Replaceable)
Yes
Data Logging
Yes
Keypad
8 electromechanical keys with graphic overlay
Software Features
Self-test and Diagnostics
Yes
Real Time Clock
Yes
Password Protection
Yes
Reset Function
Yes
Programmable Alarms
High, low, error, calibration/offline
Measurement, Active
Signal noise, temperature
Calibration
Sensor response, drift, E
0
, slope, temperature