Analyzer Operation
III-30
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
Default
Operator Action
Scrolling Marquee
Main Display
Notes
TEST (cont’d)
To Test SOUt:
4-20
Press
/
to scroll between
4-20 and rLY settings and select 4-20
SELECT mA OR RELAY
tESt
4 - 20
(flashing)
Press
to test 4-20 outputs
SELECT TEMP OR SENSOR
OUTPUT
4 - 20
tOUt or SOUt
(flashing)
Press
/
to scroll between
SOUt and tOUt settings and select SOUt
SELECT TEMP OR SENSOR
OUTPUT
4 - 20
SOUt
(flashing)
Press
to display the SOUt low value
Actual
low value
displayed
mA SENSOR OUTPUT LOW
VALUE
LO
# . #
4.0 mA are sourced at
output terminal 9
Press
to display the SOUt high value
Actual
high value
displayed
mA SENSOR OUTPUT HIGH
VALUE
HI
#. #
20.0 mA are sourced
at output terminal 9
Press
to accept the test and return
to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Display for two
channel analyzer only
Press
to return to the channel
specific menu options in the setup mode
Action for two
channel analyzer only
TEST
(flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of
two channel analyzer
Press
to test the mA outputs (4-20)
and relays (rLY)