background image

4

The primary maintenance for the NanoDrop Lite is to keep the 
pedestal surfaces clean.

1. Pipette 3 μl of deionized water (dH

2

O) onto the bottom pedestal. 

Do not use a squirt bottle to apply dH

2

O or any other liquid to 

the surface of the instrument. 

2. Lower the arm to form a liquid column; let it sit for 

approximately 2-3 minutes.

3. Wipe away the water from both the upper and lower pedestals 

with a dry, lint-free lab wipe.

Between measurements: 

Wipe the sample from both the upper and 

lower pedestals with a clean, dry, lint-free lab wipe, to prevent sample 
carryover and avoid residue buildup.

Between users: 

A final cleaning of both measurement surfaces with 

dH

2

O is recommended after the last sample measurement.

See the

 Pedestal Cleaning and Reconditioning

 guide for more details.

Trademark Information

Trademarks are the property of Thermo Fisher Scientific Inc. and its 
subsidiaries.

Contact Information

Thermo Fisher Scientific
3411 Silverside Road
Bancroft Building, Suite 100
Wilmington, DE 19810 U.S.A.
Telephone: (302) 479-7707
Toll free: (877) 724-7690 (US & Canada)
Fax: (302)792-7155
E-mail: [email protected]
www.thermofisher.com/nanodrop

NOTICE    

A maximum of 500 samples will be saved in the 

instrument and are available to be transferred to a USB memory 
device at any time. Sample #501 will replace sample #1.

Cleaning the Pedestals

Содержание NanoDrop Lite

Страница 1: ... customer USB memory device Memory device for customer data Keypad Operation Key or button Function When pressed initiates a specific function as displayed above each button Functions will change depending on the screen Arrow keys that control the direction of the cursor Select Accepts entered or selected values Advances to the next parameter or screen Home Returns to the main menu Blank Initiates...

Страница 2: ...ument identification Firmware Version Version number of firmware currently installed Calibration Date Date shown is the last date the calibration was checked on the NanoDrop Lite Reset Sample ID Reset Sample ID Yes No When Yes is selected the instrument will reset sample ID number back to 1 Calibration Check Verifies that the pathlengths are within tolerance It is recommended that a pedestal calib...

Страница 3: ...30 minutes when measuring many samples Figure 7 Sample measurement screen Transferring Data From Instrument to Computer These options appear when a USB device is inserted Figure 8 USB Operations The sample data is automatically saved on the instrument To transfer data from the instrument to USB memory device insert the device and then from the USB Operations menu select Save Data This file can be ...

Страница 4: ...nd avoid residue buildup Between users A final cleaning of both measurement surfaces with dH2O is recommended after the last sample measurement See the Pedestal Cleaning and Reconditioning guide for more details Trademark Information Trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries Contact Information Thermo Fisher Scientific 3411 Silverside Road Bancroft Building S...

Отзывы: