Routine Operation
Instrument options
60
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Use the
Left
and
Right
arrow keys to select whether the buzzer (audible
sound) is On or Off.
Press
OK
.
You can calibrate the dispensing valves for different liquids. Calibration
of the Multidrop Combi nL instrument is routinely carried out with the
PC software. For more information, refer to the
FILLit Software for
Multidrop Combi nL User Manual
(Cat. no. N17562).
For more information on calibrations, refer also to “Calibration” on
page 27 and “Dispensing volume and liquid” on page 36.
Liquid factor
is a step where
dispensing accuracy
can be verified and
corrected at one selected volume.
Factor
is a correction factor which can be used, for example, after speed
(pressure) or liquid change, where there is no calibration curve for a
new speed or a new liquid.
First, dispensing of multiple dispenses into a microplate strip is carried
out according to the volume selected in the Main >
Volume and
Liquid
window. Then a gravimetric measurement of the strip is
performed. After that the results are entered into the formula and a
new liquid factor is calculated. Refer to “Accuracy check” on page 61
and “Factor calculation” on page 63.
Go to the
Options
menu. Select the
Liquid factor
row.
Press
OK
.
Calibration
Liquid factor (standalone
calibration)
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...