Maintenance
Regular and preventive maintenance
Thermo Fisher Scientific
Thermo Scientific Multidrop Combi nL User Manual
71
Detaching the dispensing valve tube ends is only recommended if the
basic cleaning procedures, tip washing and emptying are not considered
sufficient.
2.
Detach the flexible dispensing valve tube ends (Figure 5–25).
Figure 5–25.
Detaching a dispensing valve tube end
3.
Place the tube ends into a waste container, for example, a decanter.
4.
Add water or detergent, for example 1% Micro-90, into the reagent
reservoir.
5.
Press the
PRIME
button to raise the pressure. This action leads to
fluid dripping into the waste container from the valve tube ends.
6.
When you have rinsed enough, release the pressure by pressing the
PRESSURE
button or by pushing the protective cover in.
7.
Repeat Steps 3 to 5 with the fluids you want to use, if necessary.
8.
Prime with distilled and filtered laboratory-grade water.
9.
Reattach the 43 µm reagent filter into the reagent reservoir (replace,
if necessary) by pushing the reagent filter into the cap tube end.
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...