Routine Operation
Protocol settings
Thermo Fisher Scientific
Thermo Scientific Multidrop Combi nL User Manual
45
Use the
Right
and
Left
arrow keys to select the speed of the shake. The
available speed settings are
Low
,
Med
(Medium) and
High
. The default
speed is
Low
.
The speed settings with regard to the movement distance in millimeters
(mm) and the frequency (cycles per second) in hertz (Hz) are presented
in Table 4–5.
Table 4–5.
Speed settings vs. distance movement and frequency
Speed
Movement distance (mm)
Frequency (Hz)
Low
2
10
Medium
1
15
High
1
20
Press
OK
.
This main view row includes both the dispensing height and the
dispensing offset functions.
Go to the
Settings
menu. Select the
Offsets
row.
Press
OK
.
The track mechanism and the dispensing valve head move to the setting
position. When you adjust the height, the dispensing valve head moves
next to the plate. On the other hand, when you adjust the X/Y offset,
the dispensing valve head moves over the first column of the plate, but
the default is always 1 mm above the plate. Adjust the optimal offset
values visually.
Dispensing offsets
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...