Routine Operation
Dispensing parameters
36
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Go to the
Main
menu. Select the
Volume and Liquid
row.
Press
OK
. The default dispensing volume for the selected plate type is
shown.
Select the
dispensing volume
(
Volume
) using the
Up
and
Down
arrow
keys. You can speed up the selection by holding down the arrow key
continuously.
Then select
Liquid
(Calibration) using the
Left
and
Right
arrow keys.
The different liquids set are:
l
Water
l
1% BSA (bovine serum albumin)
l
30% glycerol
l
DMSO (dimethyl sulfoxide)
l
Methanol
If you have a different liquid than the five liquids mentioned above, you
can calibrate the new liquid with FILLit Software or use the liquid
factor to calibrate the set volume.
The optimal speed, dispensing volume range and the liquid factor for
the dispensed liquid are displayed in the
Volume and Liquid
window.
Dispensing volume
and liquid
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...