Thermo Fisher Scientific
Operation Manual HM335S
59
Connect the integrated light to the power outlet with the power cord [b]
and turn it and off with the switch [c].
C
AUTION
!
Before using the large field magnifier for the first time, please check if
the voltage conditions at the installation site comply with the power
requirements and frequency noted on the supply unit of the magnifier.
Please use the enclosed fastening elements to attach the magnifier to the
microtome.
• Remove the cover plate on the upper side of the microtome hood.
• Remove the front plastic caps from the upper side of the housing.
• Carefully introduce the four sleeves [g] into the drilled holes with
guidance of the screws [e].
C
AUTION
!
Please be sure that the sleeves do not fall into the interior of the micro-
tome!
Note!
Make sure the plate [f] is mounted in such a way that the peg [d] is closer to
the front of the microtome!
• Put the plate [f] on the sleeves and fasten it to the microtome
with four screws [e].
• Put the movable plastic socket, which is located at the lower side of
the magnifier, on the peg [d].
• The magnifier can be adjusted forwards and backwards for the most
comfortable viewing position.
Note!
If the large field magnifier is not used, slightly raise it and turn it sideways
on the peg.
Installing the Large Field
Magnifier on the Microtone
Operating Instructions
Large Field Magnifier
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com