With this option, you return all settings to factory defaults.
Microm HM355S
Factory Defaults
Mode
Cont.
Auto-V-Reduction
100
Retraction
Off
Language
German
Fine
0,5 µm
Trim
5 µm
Speed
0
Note!
By selecting this function, language is reset to german!
Note!
The memory function is used to return to the same position as for first cuts.
This function can only be used for the setting with which blocks are cut,
which have been embedded in the same molds. The blocks must be of similar
height.
• To set the so-called “first-cut-position“, move the specimen clam-
ping forward with the coarse feed knob [B] until the specimen is
positioned close to the knife edge.
• To store this position, press the button MEMO [6] for approx. 1 sec.
“Pos. stored“ is then briefly shown on the display.
• Then proceed with work (trimming, first-cuts, fine sectioning).
When the work on the block has been finished, briefly press the
MEMO button. The specimen clamping then moves backwards to
unclamp the specimen and insert a new specimen.
The specimen surface is now in the “first-cut-position“.
W
ARNING
!
The stored cutting position can only be used effectively when blocks
with the same height are cut. Neither adjustments on the knife carrier
nor the knife carrier must be moved on the consoles.
W
ARNING
!
When moving the knife carrier, a new first-cut-position must be selected.
Otherwise the danger of a collision with injuries might arise.
When turning on the instrument again later, the first-cut-position must
be selected and stored again for safety reasons.