Thermo Fisher Scientific
Operation Manual HM335S
45
Note!
Frequent use of the clamping lever can result in loss of optimal clamping
strength. If the necessary readjustment is not carried out, it might be possible
that the specimen clamp does not clamp anymore.
The clamping lever [a] should be in an almost upright position.
• To determine the clamping position of the clamping lever [a], adjust
the inner screw [c] on the backside of the orienting adapter [d]
using an allen key (size 3 mm).
• Turn the allen key in a clockwise direction if no clamping is possible
or if the position of the clamping lever [a] is too high.
• If the clamping position of the clamping lever [a] is too low, turn
the allen key in a counter-clockwise direction.
• Insert the specimen clamp again and check the clamping position.
If further adjustment is necessary, please perform above described
procedure again.
Re-adjusting Specimen
Clamps
Operating Instructions
Adapters for Specimen Clamping
[c]
[d]