Thermo Fisher Scientific
Operation Manual HM335S
51
The disposable blade carrier ER is designed to take all commercially
available high and low profile blades.
• Insert the blade into the slot behind the clamping plate [e] by tur-
ning the clamping lever [g] towards the front. Swing the knife guard
[d] to the front.
• Loosen the clamping lever [g] and, if necessary, slightly push the
lower part of the clamping plate [e] as well.
• Insert the blade on the rail [c] and push it from the side to the
middle.
• Afterwards, return the clamping lever [g] upright, thus locking
the blade in position.
• After loosening the clamping lever [b] and after having moved
the knife guard [d] upwards, move the blade together with the entire
upper part without having to loosen the blade clamping.
• This way, the entire cutting length of the blade can be used.
• Then tighten the clamping lever [b].
The lever [b] can be removed by pulling it off towards the side.
Disposable Blade
Carrier “ER”
Using Low Profile Blades
[b]
[d]
[a]
[k]
[e]
[i]
[c]
[g]
[h]
[f]
Operating Instructions
Knife and Blade Carriers
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com