5
Acquiring ESI Sample Data Using the Tune Plus Window
Setting Up to Acquire MS/MS Data in the Full Scan Type
Thermo Electron Corporation
LTQ XL Getting Started
85
Figure 37.
Define Scan dialog box, showing initial settings to optimize the isolation width of an MS/MS experiment for
reserpine
b. Verify that the values in your dialog box are the same as those shown
. Start with a relatively wide Isolation Width. Leave the
Define Scan dialog box open.
4. At this time you might want to turn on your LC pump and specify a
flow rate of 0.4 mL/min, for example, to ensure that your system does
not leak.
5. Click the
Syringe Pump
button to display the Syringe Pump dialog
box. See
6. Turn on the syringe pump and set an infusion flow rate of 5
μ
L/min, as
follows:
a. In the Flow Control group box, click the
On
option button to make
active the Flow Rate spin box.
b. Type
5
in the Flow Rate spin box to specify a rate of 5
μ
L/min.
Note
This procedure assumes that you are using the 500-
μ
L
Unimetrics syringe that is provided with your LTQ XL system. If
you are using another type of syringe, select the option button
corresponding to your syringe.