Parameter Functions 

CAUTION! 

Observe the following precautions 

when operating the oven: 

Wear protective 
Wear protective gloves. 

Use tongs to insert and remove oven 

Do  not allow the load to touch the oven walls. 

CAUTION! 

Do  not attempt to operate the oven 

above the maximum rated temperature of 

260°C. 

Operation above 

260°C 

can cause damage to the 

oven. 

4.1 

Oven Start Up 

To start the oven, complete the following steps: 

1. Turn on main power switch. Wait until the controller runs 

initial diagnostics. 

2. Verify the vacuum pump connection (see Section 3.3 on 

3. Select the desired setpoints (refer to Section 5.3 on page 6 if 

you have a 

controller, Section 

6 if you have a UP150 

controller). 

5  Operation 

Controller 

The oven temperature controller is configured and tuned at the 
factory to function well for most applications. Occasionally, it 

may be advisable to configure the temperature controller 
differently to suit a particular working environment or process. 

Table 

3. 

Pushbutton  Keypad 

Before reconfiguring the controller, read 

this chapter and the 

operation manual. 

Reconfiguring the controller can change the unit 

SC 

dr 

and design 

which can 

hamper performance and make the equipment 

dangerous to use. 

Super Control 

action 

This chapter provides brief instructions on how to perform the 
following configuration changes: 

Setting the temperature 
Setting the Overtemperature Protection Temperature 
Changing between 

and 

Setting the ramp to 

time 

Autotuning the controller 

Detailed instructions on configuring the temperature controller 
are found in the 

operation manual. 

5.1 

Normal Controller Operation 

The Temperature Controller senses the chamber air temperature 
of the oven (the 

PV, or process value) and supplies the heat 

necessary to achieve the desired setpoint. The controller includes 
an LED display and a pushbutton keypad. Refer to Table 2 and 
Table 3 for lists of displayed parameters and keypad functions. 

Button 

Function 

Pressing and holding the SETIENT for three seconds 
advances the display to the Operation Parameters 
Menu. 
While in the Operation Parameters Menu, use 
SETIENT to move from one parameter to the next, and 
to register changes you have made in 

and 

parameter values. 
Holding SETIENT for three seconds exits either the 
Operation or Setup Parameters menu. 

Use the Up Arrow button to increase the temperature 

display and to change parameter values in the 

Operation and Setup Parameter menus. Whenever 
you change the value of a 

or parameter, the 

decimal point flashes to remind you to register the 

changed value with SETIENT. 

Use the Down Arrow button to decrease the 
temperature 

display and to change parameter 

values in the Operation and Setup Parameter menus. 
Whenever you change the value of a 

or 

parameter, the decimal point flashes to remind you to 
register the changed value with SETIENT. 

Содержание Lindberg/Blue M VO Series

Страница 1: ...f Thermo Fisher Scientific LindbergIBlue M Vacuum Ovens V0914 I V01218 I V01824 Installation and Operating Manual 34598H19 E 11 2011 Part of Thermo Fisher Scientific Therrno S C I E N T I F I C...

Страница 2: ...Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any par...

Страница 3: ......

Страница 4: ...grammingOperation Entering a Program 9 6 3 1 Entering ProgrammingMode 9 6 3 2 Entering Program Parameters 10 6 3 3 RunningaProgram 11 6 3 4 EndingaProgram 11 6 3 5 ChangingaProgram 11 6 4 Auto Tuning...

Страница 5: ...electronic control models with suffix P have programmable controls Built in overtemperature protection Fully flexible vacuudpurge release system Vacuum system includes inert gas injection valve fresh...

Страница 6: ...y damage can result from processing combustible volatile fluids or materials which emit explosivevapors Voltage 120 VAC 50160Hz Single Phase 240 VAC 50160 H Single Phase 120 VAC 50160 Hz Single Phase...

Страница 7: ...I 1 1 1 Power cord Overtemperatureresetbutton WARNING When installing maintaining or removing the refractory insulation the following precautions will minimize airborne dust and fiber Keeppersonnel n...

Страница 8: ...he front panel display reads gauge pressure the value is relative to the local atmospheric pressure Use in a location with reduced atmospheric pressure will result in the gauge displaying diminished v...

Страница 9: ...nstructions on how to perform the following configuration changes Setting the temperature Setting the OvertemperatureProtection Temperature Changing between celsius and fahrenheit Setting the ramp to...

Страница 10: ...and hold SETIENT for 3 seconds to enter the Operating Parameters Menu and display the A1 parameter 10 Usethe Abutton to set the A1 parameter to the desired overtemperature limit in O F 11 Press and re...

Страница 11: ...ter 5 Press and release the arrow buttons to show on in the lower display 6 Press SETIENT once to enter the auto tune mode and exit the Operating Parameters menu The controller will cycle three times...

Страница 12: ...ler includes an automatic tuning feature and Super control to improve the performance Refer to the UP150 manual for detailed information OFF OFF Pid OFF 32 4 1384 346 1 OFF 0 0 0 ProgramHold Segment A...

Страница 13: ...is display mode The SETJENT button will register setpoint value changes until these values are changed again 6 2 3 Exiting Local Mode To exit Single Setpoint or Local Mode and turn off the energy to t...

Страница 14: ...to the tM3 display If this value is correct and not changed press and release the SETENT button once to advance to the tM3 display 8 tM3 represents the third time period for the unit to reach the targ...

Страница 15: ...time length needed to get to the target setpoint SP1 7 Press and release SETENT button to display other setpoints and segment time lengths Use the arrow buttons to change the temperature setpoints and...

Страница 16: ...nstead of the factory setting of C or from O F to C follow these steps These changes will alter the controller input type and associated scale dependant parameters AND ERASE the storedprogram to defau...

Страница 17: ...f properly WARNING Disconnect oven from main power A before attemptingany maintenance to oven or its controls Regular maintenance is required to keep the ovenrunning at optimumlevels 1 Clean theovenin...

Страница 18: ...t pump down properly due to misaligneddoor Power to pump is off Use of extensioncord results in insufficient amperage Pump is damaged The discharge line is blocked No pump oil or lost viscosity overhe...

Страница 19: ...d C Units Door Magnet Catch LevellingFeet AluminumFull Shelf AluminumHalf Shelf 118975 380231 101 38024H01 38025H01 4 38026H01 38028H01 38028H04 Optional Gaskets 38027H01 38023H01 38024H01 38025H01 4...

Страница 20: ...KOGAWA UT150 CONTROL NOT TO SCALE ON g05008 I IRE1 Reference V0914 1218 1824 al l a dlrlmTE 5 18 99 T E R A N C E UNLESS OTKRWISE SPKIFIEO 1M VAC EbR A I U 03zn UNLESS OTHERWISE MTEO XX f 032 XXX f 01...

Страница 21: ...en finden Sie unter www thermo com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici W...

Страница 22: ...sed ___________________________ The above information can be found on the data plate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or spe...

Страница 23: ......

Страница 24: ...www thermoscientific com...

Отзывы: