Installation and Operation
Laboratory Vacuum Ovens
6
-
UP150
Table
4.
UP150 Parameter Functions
Models only)
RUN
HOLD
Table 5.
Pushbutton Keypad
Figure 2.
UP150
Control Panel
parameter
The oven temperature controller is configured and tuned at the
factory to function well for most applications. Occasionally, it
may be advisable to configure the temperature controller
differently to suit a particular working environment or process.
Before reconfiguring the controller, read
this chapter carefully.
the controller can
change the unit characteristics and design
parameters, which can hamper performance and
make the equipment dangerous to use.
Value
For more detailed instructions, refer to the Yokogawa UP150
manual.
Description
Operating parameters (access
by holding the
key)
6.1
Normal Controller Operation
At
P
I
Ct
FL
The Temperature Controller senses the chamber air temperature
of the oven (the
PV, or process value) and supplies the heat
necessary to achieve the desired setpoint. The controller includes
an
LED display and a pushbutton keypad. Refer to Table 2 and
Table 3 for lists of displayed parameters and keypad functions.
The UP150 controller will accept a single program of up to 16
segments. This controller includes an automatic tuning feature
and "Super" control to improve the performance. Refer to the
UP150 manual for detailed information.
OFF
OFF
Pid
32.4
1384
346
OFF
0.0
0
Program Hold
Segment Advance
Control mode
Auto tuning
Proportional band
Integral time
Derivative time
Heat cycle time
Sensor filter
PV bias (offset)
Key lock
Setup parameters (access by setting
In
SC
Button
a
Function
Pressing and holding the SETIENT for three seconds
advances the display to the Operation Parameters
Menu.
While in the Operation Parameters Menu, use
SETIENT to move from one parameter to the next, and
to register changes you have made in
and
parameter values.
Holding SETIENT for three seconds exits either the
Operation or Setup Parameters menu.
Use the Up Arrow button to increase the temperature
display and to change parametervalues in the
Operation and Setup Parameter menus. Whenever
you change the value of a
or parameter, the
decimal point flashes to remind you to register the
changed value with SETIENT.
While in operating mode, pressing this key stops
(resets) program operation.
Use the Down Arrow button to decrease the
temperature
display and to change parameter
values in the Operation and Setup Parameter menus.
Whenever you change the value
of a
or
parameter, the decimal point flashes to remind you to
register the changed value with SETIENT.
While in operating mode, pressing this key starts
(runs) a program
5
ON
Input type
(J thermocouple in
Super function