Installation and Operation 

Laboratory Vacuum Ovens 

5.2  Setting the Temperature 

To set the temperature to the desired setpoint, complete the 
following steps: 

1. Press 

or 

until the desired 

is indicated on the 

bottom line of the display. 

2. Press 

SETIENT 

to register the new setpoint. 

5.3 

Setting the Overtemperature Protection (OTP) 
Temperature 

The high limit alarm system with the temperature controller 
disables the heater output. To set the alarm on the temperature 
controller (typically 5°C above the desired main temperature 

setpoint), complete the following steps: 

1. Press and hold 

SETENT 

for 3 seconds, until 

is displayed 

on the upper line. 

2. Press 

or 

until the desired overtemperature limit 

shows on the bottom line of the display. 

3. Press 

SETENT 

to register the new overtemperature alarm 

setpoint. 

4. Press and hold 

SETENT 

for 3 seconds to return to the normal 

display. 

5.4 

Changing Between Celsius and Fahrenheit 

The controller is factory-set to operate with degrees Celsius. To 
change the display modes and parameter settings to the 

Fahrenheit scale, you will need to change the Input Type 
parameter 

In 

and also the values of various scale-dependent 

parameters. If during this procedure the buttons are inactive for 
more than two minutes, the controller will return to the standard 

display. 

To change from Celsius to Fahrenheit: 

With the controller operating, access the Operating 
Parameters  menu by pressing and holding 

SETIENT 

for 3 

seconds. 

2. Press and release 

SETIENT 

repeatedly until the upper display 

reads 

3. Press 

until the displayed value of 

is -1; then press 

SETIENT 

to access the Setup Parameters menu (refer to 

Table 2 on page 

5). 

4. The first setup parameter displayed is Input 

(In). Press 

to change its value from 

to 35. After making this 

adjustment (and all following parameter adjustments) be sure 
to press and release 

SETIENT 

again to register the change. 

5. Press and release 

SETIENT 

to advance to the SPH parameter 

and change its value to 500. 

6. Press and release 

SETIENT 

to advance to the SPL parameter 

and change its value to 32. 

7. Press and release 

SETENT 

to advance to the 

parameter 

and change its value to 

1. 

8. Press and hold 

SETENT 

for 3 seconds to exit the Setup 

Parameters Menu. 

9. Press and hold 

SETIENT 

for 3 seconds to enter the Operating 

Parameters Menu and display the 

parameter. 

the 

to set the 

parameter to the desired 

overtemperature limit in 

11. Press and release 

SETIENT 

to advance to the P parameter and 

change its value to 16.2. 

12. Press and hold 

SETIENT 

for 3 seconds to exit the Operating 

Parameters Menu. 

13. The new temperature units are now effective. Follow the 

instructions in Section 5.2 to reset the temperature 

in 

14. Apply the 

label over the 

label on the control panel. 

5.5 

Setting the Ramp to 

Rate 

The Ramp Rate feature allows the chamber to be heated or 
cooled at any rate slower than the maximum capability of the 
unit.  To fine tune ramp rates, you may need to test using loads 

with similar mass and thermal properties to loads you intend to 
use in oven applications. 

To set the ramp to 

time, complete the following steps. If 

during this procedure the buttons are inactive for more than two 
minutes, the controller will return to the standard display. 

1. With the controller operating, press and hold 

SETIENT 

for 

seconds to enter the Operating Parameters menu. 

2. Press and release 

SETIENT 

until the 

parameter is on the 

upper display. 

3. Press the down arrow button to show 

and press 

SETIENT 

once to enter the Setup Parameters menu. 

4. Press and release 

SETIENT 

until the Upr parameter is on the 

upper display. 

5. Press the arrow buttons to select the new Up Ramp Rate 

value, in 

per minute or 

per minute, or 

Press and 

release 

SETENT 

to register the value change. 

6. Press and release 

SETIENT 

until the 

parameter is on the 

upper display. 

7. 

Press the arrow buttons to select the new Down Ramp Rate 

value, in 

per minute or 

per minute, or 

Press and 

release 

SETIENT 

to register the value change. 

8. Press and hold 

SETENT 

for three seconds to exit the Setup 

Parameters  menu. 

9. The new Ramp Rates are now effective. 

10. Follow the instructions in Section 5.2 to reset the temperature 

setpoint. 

Note: 

The ramp rate begins when the 

button is 

pressed after the target 

is selected.  The 

display on the controller will show the changing 
at the selected ramp rate. 

To view the target 

during the ramp rate, press and release 

an arrow button. The lower display will show the 

next to 

the selected target setpoint. Return to the ramping 
display by pressing and releasing the other arrow button. 

Содержание Lindberg/Blue M VO Series

Страница 1: ...f Thermo Fisher Scientific LindbergIBlue M Vacuum Ovens V0914 I V01218 I V01824 Installation and Operating Manual 34598H19 E 11 2011 Part of Thermo Fisher Scientific Therrno S C I E N T I F I C...

Страница 2: ...Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any par...

Страница 3: ......

Страница 4: ...grammingOperation Entering a Program 9 6 3 1 Entering ProgrammingMode 9 6 3 2 Entering Program Parameters 10 6 3 3 RunningaProgram 11 6 3 4 EndingaProgram 11 6 3 5 ChangingaProgram 11 6 4 Auto Tuning...

Страница 5: ...electronic control models with suffix P have programmable controls Built in overtemperature protection Fully flexible vacuudpurge release system Vacuum system includes inert gas injection valve fresh...

Страница 6: ...y damage can result from processing combustible volatile fluids or materials which emit explosivevapors Voltage 120 VAC 50160Hz Single Phase 240 VAC 50160 H Single Phase 120 VAC 50160 Hz Single Phase...

Страница 7: ...I 1 1 1 Power cord Overtemperatureresetbutton WARNING When installing maintaining or removing the refractory insulation the following precautions will minimize airborne dust and fiber Keeppersonnel n...

Страница 8: ...he front panel display reads gauge pressure the value is relative to the local atmospheric pressure Use in a location with reduced atmospheric pressure will result in the gauge displaying diminished v...

Страница 9: ...nstructions on how to perform the following configuration changes Setting the temperature Setting the OvertemperatureProtection Temperature Changing between celsius and fahrenheit Setting the ramp to...

Страница 10: ...and hold SETIENT for 3 seconds to enter the Operating Parameters Menu and display the A1 parameter 10 Usethe Abutton to set the A1 parameter to the desired overtemperature limit in O F 11 Press and re...

Страница 11: ...ter 5 Press and release the arrow buttons to show on in the lower display 6 Press SETIENT once to enter the auto tune mode and exit the Operating Parameters menu The controller will cycle three times...

Страница 12: ...ler includes an automatic tuning feature and Super control to improve the performance Refer to the UP150 manual for detailed information OFF OFF Pid OFF 32 4 1384 346 1 OFF 0 0 0 ProgramHold Segment A...

Страница 13: ...is display mode The SETJENT button will register setpoint value changes until these values are changed again 6 2 3 Exiting Local Mode To exit Single Setpoint or Local Mode and turn off the energy to t...

Страница 14: ...to the tM3 display If this value is correct and not changed press and release the SETENT button once to advance to the tM3 display 8 tM3 represents the third time period for the unit to reach the targ...

Страница 15: ...time length needed to get to the target setpoint SP1 7 Press and release SETENT button to display other setpoints and segment time lengths Use the arrow buttons to change the temperature setpoints and...

Страница 16: ...nstead of the factory setting of C or from O F to C follow these steps These changes will alter the controller input type and associated scale dependant parameters AND ERASE the storedprogram to defau...

Страница 17: ...f properly WARNING Disconnect oven from main power A before attemptingany maintenance to oven or its controls Regular maintenance is required to keep the ovenrunning at optimumlevels 1 Clean theovenin...

Страница 18: ...t pump down properly due to misaligneddoor Power to pump is off Use of extensioncord results in insufficient amperage Pump is damaged The discharge line is blocked No pump oil or lost viscosity overhe...

Страница 19: ...d C Units Door Magnet Catch LevellingFeet AluminumFull Shelf AluminumHalf Shelf 118975 380231 101 38024H01 38025H01 4 38026H01 38028H01 38028H04 Optional Gaskets 38027H01 38023H01 38024H01 38025H01 4...

Страница 20: ...KOGAWA UT150 CONTROL NOT TO SCALE ON g05008 I IRE1 Reference V0914 1218 1824 al l a dlrlmTE 5 18 99 T E R A N C E UNLESS OTKRWISE SPKIFIEO 1M VAC EbR A I U 03zn UNLESS OTHERWISE MTEO XX f 032 XXX f 01...

Страница 21: ...en finden Sie unter www thermo com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici W...

Страница 22: ...sed ___________________________ The above information can be found on the data plate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or spe...

Страница 23: ......

Страница 24: ...www thermoscientific com...

Отзывы: