Installation and Operation 

Laboratory Vacuum Ovens 

6.2  Single 

Operation 

The following sections describe how to operate the controller in 
single 

(local) mode. Use this mode when you only need 

to run the oven with a specific 

and do not require a 

programmed sequence of steps. 

Setting High Temperature Alarm Setpoint: 

1. Press and HOLD for three seconds the 'SETENT'  button to 

display 

2. Press and release the 'SETENT'  button to display 
3. Press the 

Button to show the lower display 

value of 

"1". 

4. Press and release the 'SETENT'  button to select this new 

value and advance to the "SSP 

25" display. 

5. Press and release the 'SETENT" button until the High 

Temperature Alarm 

value is displayed as 

"Al". 

6. Select an alarm 

10°C above the target 

to be 

selected. 

7. Press and release the 

button to place this new 

value in the controller memory. 

8. Press and HOLD for three seconds the 

button to 

exit this menu. 

Accessing Local Mode 

1. Press and hold for three seconds the 

button to 

display 

2. Press and release the 'UP'  button twice to select the display 

3. Press and release the 'SETENT'  button once to select Local 

Mode. This selection causes the red indicator to illuminate 
beside 

o n  the control panel. 

4. Use the 'UP'  and 'DOWN'  buttons to select the desired 

operating temperature setpoint. 

5. Press and release the 'SETENT'  button once to register the 

value. 

6. The display will then show measured temperature in the 

upper display, the present temperature 

in the lower 

display. 

7. This display and the buttons will remain active as long power 

continues to the control module. Power interruptions will 
cause the controller to enter reset or standby mode in which 
no actions are made to operate the heating equipment. 

You  may use the arrow buttons to adjust the 

(lower) 

value to be adjusted in this display mode. The 

button 

will register 

value changes, until these values are 

changed again. 

Exiting Local Mode 

To exit Single 

or Local Mode and turn off the energy to 

the heaters: 

1. Press and hold for three seconds the 'SETENT'  button to 

display 

LCL". 

2. Press and release the 'DOWN'  button twice to select the 

display 

3. Press and release the 

button once to select the 

Reset Mode. 
This selection causes the red indicator to extinguish beside 

the display label "L" that had indicated the Local Mode. 

4. This will change the display showing the measured 

temperature in the upper display, with the lower display 
showing the Start Set Point (SSP) temperature 

of the 

program. 

6.3 

Programming Operation: Entering a Program 

This section describes how to enter a simple program that is 
designed to: 

direct the controller to ramp to a higher temperature; 
stabilize; 
ramp to a lower-temperature; 
end with 

an indefinite dwell. 

If you intend to use the program features of the controller, it is 
advisable to go through all the steps in this sample program to 
familiarize yourself with the elements of programming mode. 

Note: 

the controller buttons are 

for 

minutes, 

the controller will return to the regular operator 

Entering Programming Mode 

To access the programming menu: 

1. Make sure the indicators beside " R U N  and 

on the 

controller face are off. If either indicator is on, press and hold 
the 'SETENT'  buttonuntil the display shows 

Select 

in the lower display with the 'arrow'  buttons. Press and 

release the 'SETENT'  button once. 

2. Press the 'SETENT'  button for 3 seconds to display 

in the upper display and 

in the lower display. 

3. Press and release 'SETENT'  until 

is displayed. Make 

sure the display below 

is 

(zero). If it is not 

use 'DOWN ARROW'  to select 

and press and release 

"SETENT' button to register the change to 

4. Press and release the "SETENT'  button until 

is 

displayed. 

5. At 

display, press the 'UP ARROW'  to make the lower 

display 

"1". 

6. Press and release the 

button once to enter the 

programming  menu. 

Содержание Lindberg/Blue M VO Series

Страница 1: ...f Thermo Fisher Scientific LindbergIBlue M Vacuum Ovens V0914 I V01218 I V01824 Installation and Operating Manual 34598H19 E 11 2011 Part of Thermo Fisher Scientific Therrno S C I E N T I F I C...

Страница 2: ...Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any par...

Страница 3: ......

Страница 4: ...grammingOperation Entering a Program 9 6 3 1 Entering ProgrammingMode 9 6 3 2 Entering Program Parameters 10 6 3 3 RunningaProgram 11 6 3 4 EndingaProgram 11 6 3 5 ChangingaProgram 11 6 4 Auto Tuning...

Страница 5: ...electronic control models with suffix P have programmable controls Built in overtemperature protection Fully flexible vacuudpurge release system Vacuum system includes inert gas injection valve fresh...

Страница 6: ...y damage can result from processing combustible volatile fluids or materials which emit explosivevapors Voltage 120 VAC 50160Hz Single Phase 240 VAC 50160 H Single Phase 120 VAC 50160 Hz Single Phase...

Страница 7: ...I 1 1 1 Power cord Overtemperatureresetbutton WARNING When installing maintaining or removing the refractory insulation the following precautions will minimize airborne dust and fiber Keeppersonnel n...

Страница 8: ...he front panel display reads gauge pressure the value is relative to the local atmospheric pressure Use in a location with reduced atmospheric pressure will result in the gauge displaying diminished v...

Страница 9: ...nstructions on how to perform the following configuration changes Setting the temperature Setting the OvertemperatureProtection Temperature Changing between celsius and fahrenheit Setting the ramp to...

Страница 10: ...and hold SETIENT for 3 seconds to enter the Operating Parameters Menu and display the A1 parameter 10 Usethe Abutton to set the A1 parameter to the desired overtemperature limit in O F 11 Press and re...

Страница 11: ...ter 5 Press and release the arrow buttons to show on in the lower display 6 Press SETIENT once to enter the auto tune mode and exit the Operating Parameters menu The controller will cycle three times...

Страница 12: ...ler includes an automatic tuning feature and Super control to improve the performance Refer to the UP150 manual for detailed information OFF OFF Pid OFF 32 4 1384 346 1 OFF 0 0 0 ProgramHold Segment A...

Страница 13: ...is display mode The SETJENT button will register setpoint value changes until these values are changed again 6 2 3 Exiting Local Mode To exit Single Setpoint or Local Mode and turn off the energy to t...

Страница 14: ...to the tM3 display If this value is correct and not changed press and release the SETENT button once to advance to the tM3 display 8 tM3 represents the third time period for the unit to reach the targ...

Страница 15: ...time length needed to get to the target setpoint SP1 7 Press and release SETENT button to display other setpoints and segment time lengths Use the arrow buttons to change the temperature setpoints and...

Страница 16: ...nstead of the factory setting of C or from O F to C follow these steps These changes will alter the controller input type and associated scale dependant parameters AND ERASE the storedprogram to defau...

Страница 17: ...f properly WARNING Disconnect oven from main power A before attemptingany maintenance to oven or its controls Regular maintenance is required to keep the ovenrunning at optimumlevels 1 Clean theovenin...

Страница 18: ...t pump down properly due to misaligneddoor Power to pump is off Use of extensioncord results in insufficient amperage Pump is damaged The discharge line is blocked No pump oil or lost viscosity overhe...

Страница 19: ...d C Units Door Magnet Catch LevellingFeet AluminumFull Shelf AluminumHalf Shelf 118975 380231 101 38024H01 38025H01 4 38026H01 38028H01 38028H04 Optional Gaskets 38027H01 38023H01 38024H01 38025H01 4...

Страница 20: ...KOGAWA UT150 CONTROL NOT TO SCALE ON g05008 I IRE1 Reference V0914 1218 1824 al l a dlrlmTE 5 18 99 T E R A N C E UNLESS OTKRWISE SPKIFIEO 1M VAC EbR A I U 03zn UNLESS OTHERWISE MTEO XX f 032 XXX f 01...

Страница 21: ...en finden Sie unter www thermo com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici W...

Страница 22: ...sed ___________________________ The above information can be found on the data plate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or spe...

Страница 23: ......

Страница 24: ...www thermoscientific com...

Отзывы: