Keypad Display Menu Tree of LevelPRO Gauges
MAIN MENU
APPS
SysStatus
DETECTORS
SYS CNTRL
PHYSCL I/O
COM SETUP
I/O CAL
PASSWORDS
COMMANDS
WrmCldStr
Mod/Flt Alm
CURRENT OUTPUT #A
MEASUREMENT SOURCE
RANGE SETUP
MODE AND HOLD
ALARM ACTION
DIAGNOSTICS
CALIBRATE
CURRENT OUTPUT #A
VIEW CAL POINTS
BEGIN CALIBRATION
CURRENT OUTPUT #A
CALIBRATION POINTS
2PointCal
mA Value¦ RawValue
Min 4.000000¦
4.000
Mid 12.00000¦
12.000
Max 20.00000¦
20.000
BACK
NEXT
BACK
NEXT
BACK
NEXT
CURRENT OUTPUT #A
VIEW CAL POINTS
BEGIN CALIBRATION
CALIBRATION POINTS
ENTER
CAL POINT:
2PointCal
OR
PRESS NEXT TO START
CALIBRATION
BACK
NEXT
BACK
CHANGE
NEXT
REQUIRED min mA
ENTER
REQUIRED min mA
4.000000
OR PRESS NEXT
REQUIRED min mA
4.000000
ACTUAL RAW min mA
4.000
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
BACK
CHANGE
NEXT
BACK
FREEZE
NEXT
REQUIRED max mA
ENTER
REQUIRED max mA
20.00000
OR PRESS NEXT
REQUIRED max mA
20.00000
ACTUAL RAW max mA
20.000
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
BACK
CHANGE
NEXT
BACK
FREEZE
NEXT
CURRENT OUTPUT #A
ACCEPT CAL?
BACK
YES
NO
PHYSICAL I/O
CURRENT/Vdc INPUT
DIGITAL INPUT
CURRENT OUTPUT
RELAY OUTPUT
RELAY OUTPUT
RELAY OUTPUT #A
RELAY OUTPUT #B
RELAY OUTPUT #A
OUTPUT
PROCESS FUNCTION
PULSE WIDTH
(30-200mS):
30
INVERT OUTPUT:
NO
PRESENT VALUE:
OFF
HOLD/LIVE:
LIVE
HOLD VALUE:
NONE
STATUS:
LIVE
BACK
NEXT
BACK
NEXT
BACK
NEXT
BACK
CHANGE
NEXT
RELAY OUTPUT #A
OUTPUT
PROCESS FUNCTION
PROCESS FUNCTION
PrcsAlm: Pr
ALARM ACTION
DoNothin:D
TOTAL SELECTION
NONE
DET ID:
DET 1
BACK
NEXT
BACK
CHANGE
NEXT
Figure A-9.
Physical Inputs & Outputs; Digital Input, Current Output & Relay Output (Sheet 2)
Thermo Scientific
LevelPRO User Manual
A-13